Kropff J J Bruttomesso D D Doll W W Farret A A Galasso S S Luijf Y M YM Mader J K JK Place J J Boscari F F Pieber T R TR Renard E E DeVries J H JH
Diabetes, obesity & metabolism 20140910 4
<h4>Aims</h4>To assess the accuracy and reliability of the two most widely used continuous glucose monitoring (CGM) systems.<h4>Methods</h4>We studied the Dexcom®G4 Platinum (DG4P; Dexcom, San Diego, CA, USA) and Medtronic Paradigm Veo Enlite system (ENL; Medtronic, Northridge, CA, USA) CGM systems, in 24 patients with type 1 diabetes. The CGM systems were tested during 6-day home use and a nested 6-h clinical research centre (CRC) visit. During the CRC visit, frequent venous blood glucose sampl ...[more]