Unknown

Dataset Information

0

The Impact of Different Levels of Adaptive Iterative Dose Reduction 3D on Image Quality of 320-Row Coronary CT Angiography: A Clinical Trial.


ABSTRACT:

Purpose

The aim of this study was the systematic image quality evaluation of coronary CT angiography (CTA), reconstructed with the 3 different levels of adaptive iterative dose reduction (AIDR 3D) and compared to filtered back projection (FBP) with quantum denoising software (QDS).

Methods

Standard-dose CTA raw data of 30 patients with mean radiation dose of 3.2 ± 2.6 mSv were reconstructed using AIDR 3D mild, standard, strong and compared to FBP/QDS. Objective image quality comparison (signal, noise, signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), contour sharpness) was performed using 21 measurement points per patient, including measurements in each coronary artery from proximal to distal.

Results

Objective image quality parameters improved with increasing levels of AIDR 3D. Noise was lowest in AIDR 3D strong (p ? 0.001 at 20/21 measurement points; compared with FBP/QDS). Signal and contour sharpness analysis showed no significant difference between the reconstruction algorithms for most measurement points. Best coronary SNR and CNR were achieved with AIDR 3D strong. No loss of SNR or CNR in distal segments was seen with AIDR 3D as compared to FBP.

Conclusions

On standard-dose coronary CTA images, AIDR 3D strong showed higher objective image quality than FBP/QDS without reducing contour sharpness.

Trial registration

Clinicaltrials.gov NCT00967876.

SUBMITTER: Feger S 

PROVIDER: S-EPMC4422621 | biostudies-literature | 2015

REPOSITORIES: biostudies-literature

altmetric image

Publications

The Impact of Different Levels of Adaptive Iterative Dose Reduction 3D on Image Quality of 320-Row Coronary CT Angiography: A Clinical Trial.

Feger Sarah S   Rief Matthias M   Zimmermann Elke E   Martus Peter P   Schuijf Joanne Désirée JD   Blobel Jörg J   Richter Felicitas F   Dewey Marc M  

PloS one 20150506 5


<h4>Purpose</h4>The aim of this study was the systematic image quality evaluation of coronary CT angiography (CTA), reconstructed with the 3 different levels of adaptive iterative dose reduction (AIDR 3D) and compared to filtered back projection (FBP) with quantum denoising software (QDS).<h4>Methods</h4>Standard-dose CTA raw data of 30 patients with mean radiation dose of 3.2 ± 2.6 mSv were reconstructed using AIDR 3D mild, standard, strong and compared to FBP/QDS. Objective image quality compa  ...[more]

Similar Datasets

| S-EPMC6290475 | biostudies-literature
| S-EPMC7458859 | biostudies-literature
| S-EPMC2826696 | biostudies-other
| S-EPMC3606544 | biostudies-literature
| S-EPMC5384491 | biostudies-literature
| S-EPMC8506747 | biostudies-literature
| S-EPMC7458582 | biostudies-literature
| S-EPMC3807656 | biostudies-literature
| S-EPMC5849773 | biostudies-literature
| S-EPMC4143266 | biostudies-literature