Influence of accelerometer type and placement on physical activity energy expenditure prediction in manual wheelchair users.


ABSTRACT:

Purpose

To assess the validity of two accelerometer devices, at two different anatomical locations, for the prediction of physical activity energy expenditure (PAEE) in manual wheelchair users (MWUs).

Methods

Seventeen MWUs (36 ± 10 yrs, 72 ± 11 kg) completed ten activities; resting, folding clothes, propulsion on a 1% gradient (3,4,5,6 and 7 km·hr-1) and propulsion at 4km·hr-1 (with an additional 8% body mass, 2% and 3% gradient) on a motorised wheelchair treadmill. GT3X+ and GENEActiv accelerometers were worn on the right wrist (W) and upper arm (UA). Linear regression analysis was conducted between outputs from each accelerometer and criterion PAEE, measured using indirect calorimetry. Subsequent error statistics were calculated for the derived regression equations for all four device/location combinations, using a leave-one-out cross-validation analysis.

Results

Accelerometer outputs at each anatomical location were significantly (p < .01) associated with PAEE (GT3X+-UA; r = 0.68 and GT3X+-W; r = 0.82. GENEActiv-UA; r = 0.87 and GENEActiv-W; r = 0.88). Mean ± SD PAEE estimation errors for all activities combined were 15 ± 45%, 14 ± 50%, 3 ± 25% and 4 ± 26% for GT3X+-UA, GT3X+-W, GENEActiv-UA and GENEActiv-W, respectively. Absolute PAEE estimation errors for devices varied, 19 to 66% for GT3X+-UA, 17 to 122% for GT3X+-W, 15 to 26% for GENEActiv-UA and from 17.0 to 32% for the GENEActiv-W.

Conclusion

The results indicate that the GENEActiv device worn on either the upper arm or wrist provides the most valid prediction of PAEE in MWUs. Variation in error statistics between the two devices is a result of inherent differences in internal components, on-board filtering processes and outputs of each device.

SUBMITTER: Nightingale TE 

PROVIDER: S-EPMC4425541 | biostudies-literature | 2015

REPOSITORIES: biostudies-literature

altmetric image

Publications

Influence of accelerometer type and placement on physical activity energy expenditure prediction in manual wheelchair users.

Nightingale Tom Edward TE   Walhin Jean-Philippe JP   Thompson Dylan D   Bilzon James Lee John JL  

PloS one 20150508 5


<h4>Purpose</h4>To assess the validity of two accelerometer devices, at two different anatomical locations, for the prediction of physical activity energy expenditure (PAEE) in manual wheelchair users (MWUs).<h4>Methods</h4>Seventeen MWUs (36 ± 10 yrs, 72 ± 11 kg) completed ten activities; resting, folding clothes, propulsion on a 1% gradient (3,4,5,6 and 7 km·hr-1) and propulsion at 4km·hr-1 (with an additional 8% body mass, 2% and 3% gradient) on a motorised wheelchair treadmill. GT3X+ and GEN  ...[more]

Similar Datasets

| S-EPMC5628873 | biostudies-literature
| S-EPMC5332318 | biostudies-literature
| S-EPMC4010568 | biostudies-literature
| S-EPMC4142383 | biostudies-literature
| S-EPMC5425772 | biostudies-literature
| S-EPMC6186411 | biostudies-literature
| S-EPMC9246118 | biostudies-literature
| S-EPMC7364791 | biostudies-literature
| S-EPMC8192775 | biostudies-literature
| S-EPMC3708861 | biostudies-literature