Accuracy of image guidance using free-breathing cone-beam computed tomography for stereotactic lung radiotherapy.


ABSTRACT: Movement of the target object during cone-beam computed tomography (CBCT) leads to motion blurring artifacts. The accuracy of manual image matching in image-guided radiotherapy depends on the image quality. We aimed to assess the accuracy of target position localization using free-breathing CBCT during stereotactic lung radiotherapy. The Vero4DRT linear accelerator device was used for the examinations. Reference point discrepancies between the MV X-ray beam and the CBCT system were calculated using a phantom device with a centrally mounted steel ball. The precision of manual image matching between the CBCT and the averaged intensity (AI) images restructured from four-dimensional CT (4DCT) was estimated with a respiratory motion phantom, as determined in evaluations by five independent operators. Reference point discrepancies between the MV X-ray beam and the CBCT image-guidance systems, categorized as left-right (LR), anterior-posterior (AP), and superior-inferior (SI), were 0.33 ± 0.09, 0.16 ± 0.07, and 0.05 ± 0.04 mm, respectively. The LR, AP, and SI values for residual errors from manual image matching were -0.03 ± 0.22, 0.07 ± 0.25, and -0.79 ± 0.68 mm, respectively. The accuracy of target position localization using the Vero4DRT system in our center was 1.07 ± 1.23 mm (2 SD). This study experimentally demonstrated the sufficient level of geometric accuracy using the free-breathing CBCT and the image-guidance system mounted on the Vero4DRT. However, the inter-observer variation and systematic localization error of image matching substantially affected the overall geometric accuracy. Therefore, when using the free-breathing CBCT images, careful consideration of image matching is especially important.

SUBMITTER: Kamomae T 

PROVIDER: S-EPMC4425686 | biostudies-literature | 2015

REPOSITORIES: biostudies-literature

altmetric image

Publications

Accuracy of image guidance using free-breathing cone-beam computed tomography for stereotactic lung radiotherapy.

Kamomae Takeshi T   Monzen Hajime H   Nakayama Shinichi S   Mizote Rika R   Oonishi Yuuichi Y   Kaneshige Soichiro S   Sakamoto Takashi T  

PloS one 20150508 5


Movement of the target object during cone-beam computed tomography (CBCT) leads to motion blurring artifacts. The accuracy of manual image matching in image-guided radiotherapy depends on the image quality. We aimed to assess the accuracy of target position localization using free-breathing CBCT during stereotactic lung radiotherapy. The Vero4DRT linear accelerator device was used for the examinations. Reference point discrepancies between the MV X-ray beam and the CBCT system were calculated us  ...[more]

Similar Datasets

| S-EPMC7807636 | biostudies-literature
| S-EPMC7807872 | biostudies-literature
| S-EPMC10784032 | biostudies-literature
| S-EPMC6964750 | biostudies-literature
| S-EPMC5722599 | biostudies-literature
| S-EPMC11315102 | biostudies-literature
| S-EPMC10262489 | biostudies-literature
| S-EPMC8992944 | biostudies-literature
| S-EPMC7330363 | biostudies-literature
| S-EPMC9472353 | biostudies-literature