Estimation of 24-Hour Intraocular Pressure Peak Timing and Variation Using a Contact Lens Sensor.


ABSTRACT: To compare estimates of 24-hour intraocular pressure (IOP) peak timing and variation obtained using a contact lens sensor (CLS) and using a pneumatonometer.Laboratory data collected from 30 healthy volunteers (ages, 20-66 years) in a randomized, controlled clinical trial were analyzed. Participants were housed for 24 hours in a sleep laboratory. One randomly selected right or left eye was fitted with a CLS that monitored circumferential curvature in the corneoscleral region related to the change of IOP. Electronic output signals of 30 seconds were averaged and recorded every 5 minutes. In the contralateral eye, habitual IOP measurements were taken using a pneumatonometer once every two hours. Simulated 24-hour rhythms in both eyes were determined by cosinor fitting. Simulated peak timings (acrophases) and simulated data variations (amplitudes) were compared between the paired eyes.Bilateral change patterns of average 24-hour data for the group were in parallel. The simulated peak timing in the CLS fitted eye occurred at 4:44 AM ± 210 min (mean ± SD) and the IOP peak timing in the contralateral eye at 4:11 AM ± 120 min (P=0.256, Wilcoxon signed-rank test). There was no significant correlation between the simulated data variations in the paired eyes (P=0.820, linear regression).The 24-hour CLS data showed a simulated peak timing close to the 24-hour IOP peak timing obtained using the pneumatonometer. However, the simulated variations of 24-hour data in the paired eyes were not correlated. Estimated 24-hour IOP rhythms using the two devices should not be considered interchangeable.

SUBMITTER: Liu JH 

PROVIDER: S-EPMC4468106 | biostudies-literature | 2015

REPOSITORIES: biostudies-literature

altmetric image

Publications

Estimation of 24-Hour Intraocular Pressure Peak Timing and Variation Using a Contact Lens Sensor.

Liu John H K JH   Mansouri Kaweh K   Weinreb Robert N RN  

PloS one 20150615 6


<h4>Purpose</h4>To compare estimates of 24-hour intraocular pressure (IOP) peak timing and variation obtained using a contact lens sensor (CLS) and using a pneumatonometer.<h4>Methods</h4>Laboratory data collected from 30 healthy volunteers (ages, 20-66 years) in a randomized, controlled clinical trial were analyzed. Participants were housed for 24 hours in a sleep laboratory. One randomly selected right or left eye was fitted with a CLS that monitored circumferential curvature in the corneoscle  ...[more]

Similar Datasets

| S-EPMC4671823 | biostudies-literature
| S-EPMC6136041 | biostudies-literature
| S-EPMC7987168 | biostudies-literature
| S-EPMC4420265 | biostudies-literature
| S-EPMC3522433 | biostudies-literature
| S-EPMC7053760 | biostudies-literature
| S-EPMC7795318 | biostudies-literature
| S-EPMC5907516 | biostudies-literature
| S-EPMC6026180 | biostudies-literature
| S-EPMC7683862 | biostudies-literature