ABSTRACT: Objective
To test the feasibility of a CT-compatible, ex vivo, perfused porcine heart model for myocardial perfusion CT imaging.Methods
One porcine heart was perfused according to Langendorff. Dynamic perfusion scanning was performed with a second-generation dual source CT scanner. Circulatory parameters like blood flow, aortic pressure, and heart rate were monitored throughout the experiment. Stenosis was induced in the circumflex artery, controlled by a fractional flow reserve (FFR) pressure wire. CT-derived myocardial perfusion parameters were analysed at FFR of 1 to 0.10/0.0.Results
CT images did not show major artefacts due to interference of the model setup. The pacemaker-induced heart rhythm was generally stable at 70 beats per minute. During most of the experiment, blood flow was 0.9-1.0?L/min, and arterial pressure varied between 80 and 95?mm/Hg. Blood flow decreased and arterial pressure increased by approximately 10% after inducing a stenosis with FFR ? 0.50. Dynamic perfusion scanning was possible across the range of stenosis grades. Perfusion parameters of circumflex-perfused myocardial segments were affected at increasing stenosis grades.Conclusion
An adapted Langendorff porcine heart model is feasible in a CT environment. This model provides control over physiological parameters and may allow in-depth validation of quantitative CT perfusion techniques.
SUBMITTER: Pelgrim GJ
PROVIDER: S-EPMC4491382 | biostudies-literature | 2015
REPOSITORIES: biostudies-literature
Pelgrim Gert Jan GJ Das Marco M Haberland Ulrike U Slump Cees C Handayani Astri A van Tuijl Sjoerd S Stijnen Marco M Klotz Ernst E Oudkerk Matthijs M Wildberger Joachim E JE Vliegenthart Rozemarijn R
BioMed research international 20150621
<h4>Objective</h4>To test the feasibility of a CT-compatible, ex vivo, perfused porcine heart model for myocardial perfusion CT imaging.<h4>Methods</h4>One porcine heart was perfused according to Langendorff. Dynamic perfusion scanning was performed with a second-generation dual source CT scanner. Circulatory parameters like blood flow, aortic pressure, and heart rate were monitored throughout the experiment. Stenosis was induced in the circumflex artery, controlled by a fractional flow reserve ...[more]