Noninvasive assessment of hemodynamics: a comparative analysis of fingertip pulse contour analysis and impedance cardiography.


ABSTRACT: Systemic hemodynamic assessment is useful for characterizing the underlying physiology of hypertension, selecting individualized treatment approaches, and understanding the underlying mechanisms of action of interventions. Invasive methods are not suitable for routine clinic or research use, and noninvasive methods such as impedance cardiography have technical and practical limitations. Fingertip pulse contour analysis using the Nexfin device is a novel alternative to noninvasive assessment of blood pressure and hemodynamics. Although both impedance cardiography and the Nexfin have been validated against invasive methods, the extent to which they are correlated with each other is unknown. This study is a comparative analysis of data simultaneously obtained by impedance cardiography and using the Nexfin device.As part of a larger clinical trial, 13 adults with type 2 diabetes completed cardiovascular reactivity testing on three occasions: at study baseline and after two 4-week dietary treatment periods. Blood pressure, hemodynamics, and heart rate variability were assessed at rest and during acute mental stress.Blood pressure, heart rate, and heart rate variability data were significantly correlated between the two devices, but hemodynamic data (stroke volume, cardiac output, total peripheral resistance) were not significantly correlated. Both techniques detected treatment-related changes in blood pressure and total peripheral resistance, but significantly differed in the magnitude and/or direction of treatment effects.We conclude that Nexfin is not an appropriate alternative to impedance cardiography for measurement of underlying hemodynamics in psychophysiological research, but may be useful for beat-to-beat monitoring of blood pressure and heart rate variability.

SUBMITTER: Sauder KA 

PROVIDER: S-EPMC4496278 | biostudies-literature | 2015 Aug

REPOSITORIES: biostudies-literature

altmetric image

Publications

Noninvasive assessment of hemodynamics: a comparative analysis of fingertip pulse contour analysis and impedance cardiography.

Sauder Katherine A KA   Pokorney Paige E PE   McCrea Cindy E CE   Ulbrecht Jan S JS   Kris-Etherton Penny M PM   West Sheila G SG  

Blood pressure monitoring 20150801 4


<h4>Objective</h4>Systemic hemodynamic assessment is useful for characterizing the underlying physiology of hypertension, selecting individualized treatment approaches, and understanding the underlying mechanisms of action of interventions. Invasive methods are not suitable for routine clinic or research use, and noninvasive methods such as impedance cardiography have technical and practical limitations. Fingertip pulse contour analysis using the Nexfin device is a novel alternative to noninvasi  ...[more]

Similar Datasets

| S-EPMC8517395 | biostudies-literature
| S-EPMC8805171 | biostudies-literature
| S-EPMC7299376 | biostudies-literature
| S-EPMC7539259 | biostudies-literature
| S-EPMC6153686 | biostudies-literature
| S-EPMC5797604 | biostudies-other
| S-EPMC6404458 | biostudies-literature
| S-EPMC8217257 | biostudies-literature
| S-EPMC6990978 | biostudies-literature
| S-EPMC9197037 | biostudies-literature