Improved atomic force microscopy cantilever performance by partial reflective coating.


ABSTRACT: Optical beam deflection systems are widely used in cantilever based atomic force microscopy (AFM). Most commercial cantilevers have a reflective metal coating on the detector side to increase the reflectivity in order to achieve a high signal on the photodiode. Although the reflective coating is usually much thinner than the cantilever, it can still significantly contribute to the damping of the cantilever, leading to a lower mechanical quality factor (Q-factor). In dynamic mode operation in high vacuum, a cantilever with a high Q-factor is desired in order to achieve a lower minimal detectable force. The reflective coating can also increase the low-frequency force noise. In contact mode and force spectroscopy, a cantilever with minimal low-frequency force noise is desirable. We present a study on cantilevers with a partial reflective coating on the detector side. For this study, soft (?0.01 N/m) and stiff (?28 N/m) rectangular cantilevers were used with a custom partial coating at the tip end of the cantilever. The Q-factor, the detection and the force noise of fully coated, partially coated and uncoated cantilevers are compared and force distance curves are shown. Our results show an improvement in low-frequency force noise and increased Q-factor for the partially coated cantilevers compared to fully coated ones while maintaining the same reflectivity, therefore making it possible to combine the best of both worlds.

SUBMITTER: Schumacher Z 

PROVIDER: S-EPMC4505110 | biostudies-literature | 2015

REPOSITORIES: biostudies-literature

altmetric image

Publications

Improved atomic force microscopy cantilever performance by partial reflective coating.

Schumacher Zeno Z   Miyahara Yoichi Y   Aeschimann Laure L   Grütter Peter P  

Beilstein journal of nanotechnology 20150703


Optical beam deflection systems are widely used in cantilever based atomic force microscopy (AFM). Most commercial cantilevers have a reflective metal coating on the detector side to increase the reflectivity in order to achieve a high signal on the photodiode. Although the reflective coating is usually much thinner than the cantilever, it can still significantly contribute to the damping of the cantilever, leading to a lower mechanical quality factor (Q-factor). In dynamic mode operation in hig  ...[more]

Similar Datasets

| S-EPMC7810748 | biostudies-literature
| S-EPMC7506678 | biostudies-literature
| S-EPMC5057183 | biostudies-literature
| S-EPMC1367073 | biostudies-literature
| S-EPMC8697813 | biostudies-literature
| S-EPMC3298558 | biostudies-other
| S-EPMC8793147 | biostudies-literature
| S-EPMC3869821 | biostudies-other
| S-EPMC3701428 | biostudies-literature