Comparison of four digital PCR platforms for accurate quantification of DNA copy number of a certified plasmid DNA reference material.
Ontology highlight
ABSTRACT: Digital polymerase chain reaction (dPCR) is a unique approach to measurement of the absolute copy number of target DNA without using external standards. However, the comparability of different dPCR platforms with respect to measurement of DNA copy number must be addressed before dPCR can be classified fundamentally as an absolute quantification technique. The comparability of four dPCR platforms with respect to accuracy and measurement uncertainty was investigated by using a certified plasmid reference material. Plasmid conformation was found to have a significant effect on droplet-based dPCR (QX100 and RainDrop) not shared with chip-based QuantStudio 12k or BioMark. The relative uncertainty of partition volume was determined to be 0.7%, 0.8%, 2.3% and 2.9% for BioMark, QX100, QuantStudio 12k and RainDrop, respectively. The measurements of the certified pNIM-001 plasmid made using the four dPCR platforms were corrected for partition volume and closely consistent with the certified value within the expended uncertainty. This demonstrated that the four dPCR platforms are of comparable effectiveness in quantifying DNA copy number. These findings provide an independent assessment of this method of determining DNA copy number when using different dPCR platforms and underline important factors that should be taken into consideration in the design of dPCR experiments.
SUBMITTER: Dong L
PROVIDER: S-EPMC4548226 | biostudies-literature | 2015 Aug
REPOSITORIES: biostudies-literature
ACCESS DATA