Steady pulsed imaging and labeling scheme for noninvasive perfusion imaging.


ABSTRACT:

Purpose

A steady pulsed imaging and labeling (SPIL) scheme is proposed to obtain high-resolution multislice perfusion images of mice brain using standard preclinical MRI equipment.

Theory and methods

The SPIL scheme repeats a pulsed arterial spin labeling (PASL) module together with a short mixing time to extend the temporal duration of the generated PASL bolus to the total experimental time. Multislice image acquisition takes place during the mixing times. The mixing time is also used for magnetization recovery following image acquisition. The new scheme is able to yield multislice perfusion images rapidly. The perfusion kinetic curve can be measured by a multipulsed imaging and labeling (MPIL) scheme, i.e., acquiring single-slice ASL signals before reaching steady-state in the SPIL sequence.

Results

When applying the SPIL method to normal mice, and to mice with unilateral ischemia, high-resolution multislice (five slices) CBF images could be obtained in 8 min. Perfusion data from ischemic mice showed clear CBF reductions in ischemic regions. The SPIL method was also applied to postmortem mice, showing that the method is free from magnetization transfer confounds.

Conclusion

The new SPIL scheme provides for robust measurement of CBF with multislice imaging capability in small animals.

SUBMITTER: Xu J 

PROVIDER: S-EPMC4558394 | biostudies-literature | 2016 Jan

REPOSITORIES: biostudies-literature

altmetric image

Publications

Steady pulsed imaging and labeling scheme for noninvasive perfusion imaging.

Xu Jiadi J   Qin Qin Q   Wu Dan D   Hua Jun J   Song Xiaolei X   McMahon Michael T MT   Northington Frances J FJ   Zhang Jiangyang J   Zhang Jiangyang J   van Zijl Peter C M PC   Pekar James J JJ  

Magnetic resonance in medicine 20150302 1


<h4>Purpose</h4>A steady pulsed imaging and labeling (SPIL) scheme is proposed to obtain high-resolution multislice perfusion images of mice brain using standard preclinical MRI equipment.<h4>Theory and methods</h4>The SPIL scheme repeats a pulsed arterial spin labeling (PASL) module together with a short mixing time to extend the temporal duration of the generated PASL bolus to the total experimental time. Multislice image acquisition takes place during the mixing times. The mixing time is also  ...[more]

Similar Datasets

| S-EPMC5516148 | biostudies-literature
| S-EPMC4215516 | biostudies-literature
| S-EPMC3185150 | biostudies-literature
| S-EPMC4143230 | biostudies-literature
| S-EPMC6693904 | biostudies-literature
| S-EPMC6282709 | biostudies-literature
| S-EPMC8727512 | biostudies-literature
| S-EPMC11342727 | biostudies-literature
| S-EPMC4110188 | biostudies-other
| S-EPMC4369335 | biostudies-literature