2-year follow-up report on micromotion of a short tibia stem. A prospective, randomized RSA study of 59 patients.


ABSTRACT:

Background and purpose

A shortened tibial stem could influence the early prosthetic fixation. We therefore compared the short stem to the standard-length stem using radiostereometric analysis (RSA) as primary outcome measure.

Patients and methods

60 patients were randomized to receive a cemented Triathlon total knee arthroplasty (TKA) with a tibial tray of either standard or short stem length. The patients were blinded regarding treatment allocation. The micromotion of the tibial component was measured by RSA postoperatively, at 3 months, and after 1 and 2 years; clinical outcome was measured with the American Knee Society score (AKSS) and the knee osteoarthritis and injury outcome score (KOOS).

Results

The maximum total point motion (MTPM) for the standard stem was 0.36 (SD 0.16) mm at 3 months, 0.51 (SD 0.27) mm at 1 year, and 0.54 (SD 0.28) mm at 2 years. For the short stem, it was 0.42 (0.24) mm, 0.59 (0.43) mm, and 0.61 (0.39) mm. 4 short-stemmed components and 2 standard-stemmed components had more than 0.2 mm of migration between the first- and second-year follow-up, and were classified as continuously migrating.

Interpretation

The short-stemmed cemented tibial prosthesis showed an early prosthetic migratory pattern similar to that of the standard-stemmed cemented Triathlon knee prosthesis. The number of continuously migrating tibial plates in each group is predictive of a lower revision rate than 5% at 10 years.

SUBMITTER: Molt M 

PROVIDER: S-EPMC4564782 | biostudies-literature | 2015

REPOSITORIES: biostudies-literature

altmetric image

Publications

2-year follow-up report on micromotion of a short tibia stem. A prospective, randomized RSA study of 59 patients.

Molt Mats M   Toksvig-Larsen Sören S  

Acta orthopaedica 20150101 5


<h4>Background and purpose</h4>A shortened tibial stem could influence the early prosthetic fixation. We therefore compared the short stem to the standard-length stem using radiostereometric analysis (RSA) as primary outcome measure.<h4>Patients and methods</h4>60 patients were randomized to receive a cemented Triathlon total knee arthroplasty (TKA) with a tibial tray of either standard or short stem length. The patients were blinded regarding treatment allocation. The micromotion of the tibial  ...[more]

Similar Datasets

| S-EPMC5434595 | biostudies-literature
| S-EPMC11361569 | biostudies-literature
| S-EPMC10509037 | biostudies-literature
| S-EPMC8023937 | biostudies-literature
| S-EPMC3538485 | biostudies-literature
| S-EPMC6076883 | biostudies-other
| S-EPMC6971974 | biostudies-literature
| S-EPMC4485526 | biostudies-literature
| S-EPMC3584612 | biostudies-literature
2005-01-18 | GSE1907 | GEO