Correlation between Serum Levels of 3,3',5'-Triiodothyronine and Thyroid Hormones Measured by Liquid Chromatography-Tandem Mass Spectrometry and Immunoassay.


ABSTRACT:

Objective

For measuring serum 3,3',5'-triiodothyronine (rT3) levels, radioimmunoassay (RIA) has traditionally been used owing to the lack of other reliable methods; however, it has recently become difficult to perform. Meanwhile, liquid chromatography-tandem mass spectrometry (LC-MS/MS) has recently been attracting attention as a novel alternative method in clinical chemistry. To the best of our knowledge, there are no studies to date comparing results of the quantification of human serum rT3 between LC-MS/MS and RIA. We therefore examined the feasibility of LC-MS/MS as a novel alternative method for measuring serum rT3, thyroxine (T4), and 3,5,3'-triiodothyronine (T3) levels.

Methods

Assay validation was performed by LC-MS/MS using quality control samples of rT3, T4, and T3 at 4 various concentrations which were prepared from reference compounds. Serum samples of 50 outpatients in our department were quantified both by LC-MS/MS and conventional immunoassay for rT3, T4, and T3. Correlation coefficients between the 2 measurement methods were statistically analyzed respectively.

Results

Matrix effects were not observed with our method. Intra-day and inter-day precisions were less than 10.8% and 9.6% for each analyte at each quality control level, respectively. Intra-day and inter-day accuracies were between 96.2% and 110%, and between 98.3% and 108.6%, respectively. The lower limit of quantification was 0.05 ng/mL. Strong correlations were observed between the 2 measurement methods (correlation coefficient, T4: 0.976, p < 0.001; T3: 0.912, p < 0.001; rT3: 0.928, p < 0.001).

Conclusions

Our LC-MS/MS system requires no manual cleanup operation, and the process after application of a sample is fully automated; furthermore, it was found to be highly sensitive, and superior in both precision and accuracy. The correlation between the 2 methods over a wide range of concentrations was strong. LC-MS/MS is therefore expected to become a useful tool for clinical diagnosis and research.

SUBMITTER: Sakai H 

PROVIDER: S-EPMC4591014 | biostudies-literature | 2015

REPOSITORIES: biostudies-literature

altmetric image

Publications

Correlation between Serum Levels of 3,3',5'-Triiodothyronine and Thyroid Hormones Measured by Liquid Chromatography-Tandem Mass Spectrometry and Immunoassay.

Sakai Hiroyuki H   Nagao Hidenori H   Sakurai Mamoru M   Okumura Takako T   Nagai Yoshiyuki Y   Shikuma Junpei J   Ito Rokuro R   Imazu Tetsuya T   Miwa Takashi T   Odawara Masato M  

PloS one 20151001 10


<h4>Objective</h4>For measuring serum 3,3',5'-triiodothyronine (rT3) levels, radioimmunoassay (RIA) has traditionally been used owing to the lack of other reliable methods; however, it has recently become difficult to perform. Meanwhile, liquid chromatography-tandem mass spectrometry (LC-MS/MS) has recently been attracting attention as a novel alternative method in clinical chemistry. To the best of our knowledge, there are no studies to date comparing results of the quantification of human seru  ...[more]

Similar Datasets

| S-EPMC10762560 | biostudies-literature
| S-EPMC6522465 | biostudies-other
| S-EPMC5476196 | biostudies-literature
| S-EPMC9226140 | biostudies-literature
| S-EPMC9334332 | biostudies-literature
| S-EPMC5960525 | biostudies-literature
| S-EPMC9032635 | biostudies-literature
| S-EPMC3979722 | biostudies-literature
| S-EPMC8534991 | biostudies-literature
| S-EPMC8581966 | biostudies-literature