ABSTRACT: Introduction
The purpose of this paper is to develop an easy method to generate both fat signal and banding artifact free 3D balanced Steady State Free Precession (bSSFP) images at high magnetic field.Methods
In order to suppress fat signal and bSSFP banding artifacts, two or four images were acquired with the excitation frequency of the water-selective binomial radiofrequency pulse set On Resonance or shifted by a maximum of 3/4TR. Mice and human volunteers were imaged at 7 T and 3 T, respectively to perform whole-body and musculoskeletal imaging. "Sum-Of-Square" reconstruction was performed and combined or not with parallel imaging.Results
The frequency selectivity of 1-2-3-2-1 or 1-3-3-1 binomial pulses was preserved after (3/4TR) frequency shifting. Consequently, whole body small animal 3D imaging was performed at 7 T and enabled visualization of small structures within adipose tissue like lymph nodes. In parallel, this method allowed 3D musculoskeletal imaging in humans with high spatial resolution at 3 T. The combination with parallel imaging allowed the acquisition of knee images with ~500 ?m resolution images in less than 2 min. In addition, ankles, full head coverage and legs of volunteers were imaged, demonstrating the possible application of the method also for large FOV.Conclusion
In conclusion, this robust method can be applied in small animals and humans at high magnetic fields. The high SNR and tissue contrast obtained in short acquisition times allows to prescribe bSSFP sequence for several preclinical and clinical applications.
SUBMITTER: Ribot EJ
PROVIDER: S-EPMC4591352 | biostudies-literature | 2015
REPOSITORIES: biostudies-literature
Ribot Emeline J EJ Wecker Didier D Trotier Aurélien J AJ Dallaudière Benjamin B Lefrançois William W Thiaudière Eric E Franconi Jean-Michel JM Miraux Sylvain S
PloS one 20151001 10
<h4>Introduction</h4>The purpose of this paper is to develop an easy method to generate both fat signal and banding artifact free 3D balanced Steady State Free Precession (bSSFP) images at high magnetic field.<h4>Methods</h4>In order to suppress fat signal and bSSFP banding artifacts, two or four images were acquired with the excitation frequency of the water-selective binomial radiofrequency pulse set On Resonance or shifted by a maximum of 3/4TR. Mice and human volunteers were imaged at 7 T an ...[more]