ABSTRACT: Background
Less-invasive and easy to install monitoring systems for continuous estimation of cardiac index (CI) have gained increasing interest, especially in cardiac surgery patients who often exhibit abrupt haemodynamic changes. The aim of the present study was to compare the accuracy of CI by a new semi-invasive monitoring system with transpulmonary thermodilution before and after cardiopulmonary bypass (CPB).Methods
Sixty-five patients (41 Germany, 24 Spain) scheduled for elective coronary surgery were studied before and after CPB, respectively. Measurements included CI obtained by transpulmonary thermodilution (CITPTD) and autocalibrated semi-invasive pulse contour analysis (CIPFX). Percentage changes of CI were also calculated.Results
There was only a poor cor