ABSTRACT: Background
To deliver insulin glargine 300 U/mL (Gla-300), the widely used SoloSTAR(®) pen has been modified to allow for accurate and precise delivery of required insulin units in one-third of the volume compared with insulin glargine 100 U/mL, while improving usability. Here we compare the accuracy and injection force of 3 disposable insulin pens: Gla-300 SoloSTAR(®), FlexPen(®), and KwikPen™.Methods
For the accuracy assessment, 60 of each of the 3 tested devices were used for the delivery of 3 different doses (1 U, half-maximal dose, and maximal dose), which were measured gravimetrically. For the injection force assessment, 20 pens of each of the 3 types were tested twice at half-maximal and once at maximal dose, at an injection speed of 6 U/s.Results
All tested pens met the International Organization for Standardization (ISO) requirements for dosing accuracy, with Gla-300 SoloSTAR showing the lowest between-dose variation (greatest reproducibility) at all dose levels. Mean injection force was significantly lower for Gla-300 SoloSTAR than for the other 2 pens at both half maximal and maximal doses (P < .0271).Conclusion
All tested pens were accurate according to ISO criteria, and the Gla-300 SoloSTAR pen displayed the greatest reproducibility and lowest injection force of any of the 3 tested devices.
SUBMITTER: Klonoff D
PROVIDER: S-EPMC4738217 | biostudies-literature | 2015 Aug
REPOSITORIES: biostudies-literature
Klonoff David D Nayberg Irina I Thonius Marissa M See Florian F Abdel-Tawab Mona M Erbstein Frank F Haak Thomas T
Journal of diabetes science and technology 20150826 1
<h4>Background</h4>To deliver insulin glargine 300 U/mL (Gla-300), the widely used SoloSTAR(®) pen has been modified to allow for accurate and precise delivery of required insulin units in one-third of the volume compared with insulin glargine 100 U/mL, while improving usability. Here we compare the accuracy and injection force of 3 disposable insulin pens: Gla-300 SoloSTAR(®), FlexPen(®), and KwikPen™.<h4>Methods</h4>For the accuracy assessment, 60 of each of the 3 tested devices were used for ...[more]