Coil combination for receive array spectroscopy: Are data-driven methods superior to methods using computed field maps?


ABSTRACT:

Purpose

Combining spectra from receive arrays, particularly X-nuclear spectra with low signal-to-noise ratios (SNRs), is challenging. We test whether data-driven combination methods are better than using computed coil sensitivities.

Theory

Several combination algorithms are recast into the notation of Roemer's classic formula, showing that they differ primarily in their estimation of coil receive sensitivities. This viewpoint reveals two extensions of the whitened singular-value decomposition (WSVD) algorithm, using temporal or temporal + spatial apodization to improve the coil sensitivities, and thus the combined spectral SNR.

Methods

Radiofrequency fields from an array were simulated and used to make synthetic spectra. These were combined with 10 algorithms. The combined spectra were then assessed in terms of their SNR. Validation used phantoms and cardiac (31) P spectra from five subjects at 3T.

Results

Combined spectral SNRs from simulations, phantoms, and humans showed the same trends. In phantoms, the combined SNR using computed coil sensitivities was lower than with WSVD combination whenever the WSVD SNR was >14 (or >11 with temporal apodization, or >9 with temporal + spatial apodization). These new apodized WSVD methods gave higher SNRs than other data-driven methods.

Conclusion

In the human torso, at frequencies ?49 MHz, data-driven combination is preferable to using computed coil sensitivities. Magn Reson, 2015. © 2015 The Authors. Magnetic Resonance in Medicine published by Wiley Periodicals, Inc. on behalf of International Society for Magnetic Resonance in Medicine. This is an open access article under the terms of the Creative Commons Attribution License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited. Magn Reson Med 75:473-487, 2016. © 2015 The Authors. Magnetic Resonance in Medicine published by Wiley Periodicals, Inc. on behalf of International Society for Magnetic Resonance in Medicine.

SUBMITTER: Rodgers CT 

PROVIDER: S-EPMC4744755 | biostudies-literature | 2016 Feb

REPOSITORIES: biostudies-literature

altmetric image

Publications

Coil combination for receive array spectroscopy: Are data-driven methods superior to methods using computed field maps?

Rodgers Christopher T CT   Robson Matthew D MD  

Magnetic resonance in medicine 20150328 2


<h4>Purpose</h4>Combining spectra from receive arrays, particularly X-nuclear spectra with low signal-to-noise ratios (SNRs), is challenging. We test whether data-driven combination methods are better than using computed coil sensitivities.<h4>Theory</h4>Several combination algorithms are recast into the notation of Roemer's classic formula, showing that they differ primarily in their estimation of coil receive sensitivities. This viewpoint reveals two extensions of the whitened singular-value d  ...[more]

Similar Datasets

| S-EPMC8774338 | biostudies-literature
| S-EPMC8519055 | biostudies-literature
| S-EPMC6618274 | biostudies-literature
| S-EPMC9675905 | biostudies-literature
| S-EPMC7490394 | biostudies-literature
| S-EPMC4200535 | biostudies-other
| S-EPMC3558719 | biostudies-other
| S-EPMC5828900 | biostudies-literature
| S-EPMC5537232 | biostudies-literature
| S-EPMC5728109 | biostudies-literature