ABSTRACT: Purpose
To assess the measurement variability of subsolid nodules (SSNs) in follow-up situations and to compare the degree of variability between measurement metrics.Methods
Two same-day repeat-CT scans of 69 patients (24 men and 45 women) with 69 SSNs were randomly assigned as initial or follow-up scans and were read by the same (situation 1) or different readers (situation 2). SSN size and solid portion size were measured in both situations. Measurement variability was calculated and coefficients of variation were used for comparisons.Results
Measurement variability for the longest and average diameter of SSNs was ±1.3 mm (±13.0%) and ±1.3 mm (±14.4%) in situation 1, and ±2.2 mm (±21.0%) and ±2.1 mm (±21.3%) in situation 2, respectively. For solid portion, measurement variability on lung and mediastinal windows was ±1.2 mm (±27.1%) and ±0.8 mm (±24.0%) in situation 1, and ±3.7 mm (±61.0%) and ±1.5 mm (±47.3%) in situation 2, respectively. There were no significant differences in the degree of variability between the longest and average diameters and between the lung and mediastinal window settings (p>0.05). However, measurement variability significantly increased when the follow-up and initial CT readers were different (p<0.001).Conclusions
A cutoff of ±2.2 mm can be reliably used to determine true nodule growth on follow-up CT. Solid portion measurements were not reliable in evaluating SSNs' change when readers of initial and follow-up CT were different.
SUBMITTER: Kim H
PROVIDER: S-EPMC4747473 | biostudies-literature | 2016
REPOSITORIES: biostudies-literature
Kim Hyungjin H Park Chang Min CM Song Yong Sub YS Sunwoo Leonard L Choi Ye Ra YR Kim Jung Im JI Kim Jae Hyun JH Bae Jae Seok JS Lee Jong Hyuk JH Goo Jin Mo JM
PloS one 20160209 2
<h4>Purpose</h4>To assess the measurement variability of subsolid nodules (SSNs) in follow-up situations and to compare the degree of variability between measurement metrics.<h4>Methods</h4>Two same-day repeat-CT scans of 69 patients (24 men and 45 women) with 69 SSNs were randomly assigned as initial or follow-up scans and were read by the same (situation 1) or different readers (situation 2). SSN size and solid portion size were measured in both situations. Measurement variability was calculat ...[more]