Relation between detection rate and inappropriate shocks in single versus dual chamber cardioverter-defibrillator--an analysis from the OPTION trial.
Ontology highlight
ABSTRACT: The programming of implantable cardioverter-defibrillators (ICDs) influences inappropriate shock rates. The aim of the study is to analyse rates of patients with appropriate and inappropriate shocks according to detection zones in the OPTION trial. All patients received dual chamber (DC) ICDs randomly assigned to be programmed either to single chamber (SC) or to DC settings including PARAD+ algorithm. In a post-hoc analysis, rates of patients with inappropriate and appropriate shocks were calculated for shocks triggered at heart rates ≥ 170 bpm (ventricular tachycardia zone) and at rates ≥ 200 bpm (ventricular fibrillation zone). In the SC group, higher rates of patients with total and inappropriate shocks were delivered at heart rates ≥ 170 bpm than at rates ≥ 200 bpm (total shocks: 21.1%