Suppression of skeletal muscle signal using a crusher coil: A human cardiac (31) p-MR spectroscopy study at 7 tesla.


ABSTRACT:

Purpose

The translation of sophisticated phosphorus MR spectroscopy ((31)P-MRS) protocols to 7 Tesla (T) is particularly challenged by the issue of radiofrequency (RF) heating. Legal limits on RF heating make it hard to reliably suppress signals from skeletal muscle that can contaminate human cardiac (31)P spectra at 7T. We introduce the first surface-spoiling crusher coil for human cardiac (31)P-MRS at 7T.

Methods

A planar crusher coil design was optimized with simulations and its performance was validated in phantoms. Crusher gradient pulses (100 ?s) were then applied during human cardiac (31)P-MRS at 7T.

Results

In a phantom, residual signals were 50 ± 10% with BISTRO (B1 -insensitive train to obliterate signal), and 34 ± 8% with the crusher coil. In vivo, residual signals in skeletal muscle were 49 ± 4% using BISTRO, and 24 ± 5% using the crusher coil. Meanwhile, in the interventricular septum, spectral quality and metabolite quantification did not differ significantly between BISTRO (phosphocreatine/adenosine triphosphate [PCr/ATP]?= 2.1 ± 0.4) and the crusher coil (PCr/ATP = 1.8 ± 0.4). However, the specific absorption rate (SAR) decreased from 96 ± 1% of the limit (BISTRO) to 16 ± 1% (crusher coil).

Conclusion

A crusher coil is an SAR-efficient alternative for selectively suppressing skeletal muscle during cardiac (31)P-MRS at 7T. A crusher coil allows the use of sequence modules that would have been SAR-prohibitive, without compromising skeletal muscle suppression.

SUBMITTER: Schaller B 

PROVIDER: S-EPMC4762536 | biostudies-literature | 2016 Mar

REPOSITORIES: biostudies-literature

altmetric image

Publications

Suppression of skeletal muscle signal using a crusher coil: A human cardiac (31) p-MR spectroscopy study at 7 tesla.

Schaller Benoit B   Clarke William T WT   Neubauer Stefan S   Robson Matthew D MD   Rodgers Christopher T CT  

Magnetic resonance in medicine 20150428 3


<h4>Purpose</h4>The translation of sophisticated phosphorus MR spectroscopy ((31)P-MRS) protocols to 7 Tesla (T) is particularly challenged by the issue of radiofrequency (RF) heating. Legal limits on RF heating make it hard to reliably suppress signals from skeletal muscle that can contaminate human cardiac (31)P spectra at 7T. We introduce the first surface-spoiling crusher coil for human cardiac (31)P-MRS at 7T.<h4>Methods</h4>A planar crusher coil design was optimized with simulations and it  ...[more]

Similar Datasets

| S-EPMC9297863 | biostudies-literature
| S-EPMC5697655 | biostudies-literature
| S-EPMC4347229 | biostudies-literature
| S-EPMC4500638 | biostudies-literature
| S-EPMC4106879 | biostudies-literature
| S-EPMC5828900 | biostudies-literature
| S-EPMC5076794 | biostudies-literature
| S-EPMC11190064 | biostudies-literature
| S-EPMC8590093 | biostudies-literature
| S-EPMC10931074 | biostudies-literature