Unknown

Dataset Information

0

A Clinical and Experimental Comparison of Time of Flight PET/MRI and PET/CT Systems.


ABSTRACT: PURPOSE:The purpose of the study was to compare image quality and quantitative accuracy of positron emission tomography/magnetic resonance imaging (PET/MRI) and PET/computed tomography (PET/CT) systems with time of flight PET gantries, using phantom and clinical studies. PROCEDURES:Identical phantom experiments were performed on both systems. Calibration, uniformity, and standardized uptake value (SUV) recovery were measured. A clinical PET/CT versus PET/MRI comparison was performed using [(18)F]fluoromethylcholine ([(18)F]FCH). RESULTS:Calibration accuracy and image uniformity were comparable between systems. SUV recovery met EANM/EARL requirements on both scanners. Thirty-four lesions with comparable PET image quality were identified. Lesional SUVmax differences of 4 ± 26% between PET/MRI and PET/CT data were observed (R (2) = 0.79, slope = 1.02). In healthy tissues, PET/MRI-derived SUVs were 16 ± 11% lower than on PET/CT (R (2) = 0.98, slope = 0.86). CONCLUSION:PET/MRI and PET/CT showed comparable performance with respect to calibration accuracy, image uniformity, and SUV recovery. [(18)F]FCH uptake values for both healthy tissues and lesions corresponded reasonably well between MR- and CT-based systems, but only in regions free of MR-based attenuation artifacts.

SUBMITTER: Oprea-Lager DE 

PROVIDER: S-EPMC4768240 | biostudies-literature |

REPOSITORIES: biostudies-literature

Similar Datasets

| S-EPMC7850238 | biostudies-literature
| S-EPMC4288516 | biostudies-literature
| S-EPMC7469956 | biostudies-literature
| S-EPMC6283203 | biostudies-literature
| S-EPMC9139574 | biostudies-literature
| S-EPMC5981153 | biostudies-literature
| S-EPMC6637922 | biostudies-literature
| S-EPMC6656738 | biostudies-literature
| S-EPMC6292629 | biostudies-literature
| S-EPMC7465843 | biostudies-literature