Orthogonal Injection Ion Funnel Interface Providing Enhanced Performance for Selected Reaction Monitoring-Triple Quadrupole Mass Spectrometry.


ABSTRACT: The electrodynamic ion funnel facilitates efficient focusing and transfer of charged particles in the higher-pressure regions (e.g., ion source interfaces) of mass spectrometers, thus providing increased sensitivity. An "off-axis" ion funnel design has been developed to reduce the source contamination and interferences from, e.g. ESI droplet residue and other poorly focused neutral or charged particles with very high mass-to-charge ratios. In this study, a dual ion funnel interface consisting of an orthogonal higher pressure electrodynamic ion funnel (HPIF) and an ion funnel trap combined with a triple quadrupole mass spectrometer was developed and characterized. An orthogonal ion injection inlet and a repeller plate electrode was used to direct ions to an ion funnel HPIF at a pressure of 9-10 Torr. Key factors for the HPIF performance characterized included the effects of RF amplitude, the DC gradient, and operating pressure. Compared to the triple quadrupole standard interface more than 4-fold improvement in the limit of detection for the direct quantitative MS analysis of low abundance peptides was observed. The sensitivity enhancement in liquid chromatography selected reaction monitoring (LC-SRM) analyses of low-abundance peptides spiked into a highly complex mixture was also compared with that obtained using both a commercial S-lens interface and an in-line dual-ion funnel interface.

SUBMITTER: Chen TC 

PROVIDER: S-EPMC4776758 | biostudies-literature | 2015 Jul

REPOSITORIES: biostudies-literature

altmetric image

Publications

Orthogonal Injection Ion Funnel Interface Providing Enhanced Performance for Selected Reaction Monitoring-Triple Quadrupole Mass Spectrometry.

Chen Tsung-Chi TC   Fillmore Thomas L TL   Prost Spencer A SA   Moore Ronald J RJ   Ibrahim Yehia M YM   Smith Richard D RD  

Analytical chemistry 20150708 14


The electrodynamic ion funnel facilitates efficient focusing and transfer of charged particles in the higher-pressure regions (e.g., ion source interfaces) of mass spectrometers, thus providing increased sensitivity. An "off-axis" ion funnel design has been developed to reduce the source contamination and interferences from, e.g. ESI droplet residue and other poorly focused neutral or charged particles with very high mass-to-charge ratios. In this study, a dual ion funnel interface consisting of  ...[more]

Similar Datasets

| S-EPMC2516353 | biostudies-literature
| S-EPMC4287841 | biostudies-other
| S-EPMC8762722 | biostudies-literature
| S-EPMC2724058 | biostudies-literature
| S-EPMC6002426 | biostudies-literature
| S-EPMC2646368 | biostudies-other
| S-EPMC6470114 | biostudies-literature
| S-EPMC6526401 | biostudies-literature
| S-EPMC6027907 | biostudies-literature
| S-EPMC4034390 | biostudies-literature