A novel method of creation of a fenestration in nitinol occluder devices used in closure of hypertensive patent arterial ducts.


ABSTRACT: Test occlusion with a balloon is done to predict operability of large hypertensive patent ductus arteriosus (PDA). If the fall in the pulmonary artery pressures is inadequate, a complete closure is not desired. To create a predictable premeasured fenestration in a nitinol occluder device used for closing hypertensive PDA. A large nitinol occluder device was punctured with an 18G needle to advance a 0.035? stiff guide wire through the occluder before loading it into the delivery system. The occluder with the guidewire was then deployed across the PDA. A coronary guide catheter was later threaded through the guidewire into the fabric of the device, which was still held by the delivery cable. A coronary stent was deployed across the fenestration in the occluder to keep it patent. An 8-year-old boy with Down syndrome and hypertensive PDA was hemodynamically assessed. Even though there was a fall in the pulmonary vascular resistance index and pressures on test occlusion, the pulmonary artery pressures were labile with fluctuations. A customized fenestration was made in a 16 mm muscular ventricular septal defect occluder (MVSO) with a 4.5 mm bare-metal coronary stent. The pulmonary artery pressures remained at half of the aortic pressures after the procedure. This fenestration model precisely and predictably fenestrated a large occluder device used to close a hypertensive large PDA. Long-term patency of these fenestrations has to be assessed on the follow-up, and may be improved through larger fenestrations, systemic anticoagulation and use of covered stents.

SUBMITTER: Singhi AK 

PROVIDER: S-EPMC4782470 | biostudies-literature | 2016 Jan-Apr

REPOSITORIES: biostudies-literature

altmetric image

Publications

A novel method of creation of a fenestration in nitinol occluder devices used in closure of hypertensive patent arterial ducts.

Singhi Anil Kumar AK   Sivakumar Kothandam K  

Annals of pediatric cardiology 20160101 1


Test occlusion with a balloon is done to predict operability of large hypertensive patent ductus arteriosus (PDA). If the fall in the pulmonary artery pressures is inadequate, a complete closure is not desired. To create a predictable premeasured fenestration in a nitinol occluder device used for closing hypertensive PDA. A large nitinol occluder device was punctured with an 18G needle to advance a 0.035˝ stiff guide wire through the occluder before loading it into the delivery system. The occlu  ...[more]

Similar Datasets

| S-EPMC4099166 | biostudies-literature
| S-EPMC8311012 | biostudies-literature
| S-EPMC3860766 | biostudies-literature
| S-EPMC8099582 | biostudies-literature
| S-EPMC7754477 | biostudies-literature
| S-EPMC5508309 | biostudies-literature
| S-EPMC8373728 | biostudies-literature
| S-EPMC6374772 | biostudies-literature
| S-EPMC7805018 | biostudies-literature
| S-EPMC8270784 | biostudies-literature