An Analysis of Plan Robustness for Esophageal Tumors: Comparing Volumetric Modulated Arc Therapy Plans and Spot Scanning Proton Planning.


ABSTRACT:

Purpose

Planning studies to compare x-ray and proton techniques and to select the most suitable technique for each patient have been hampered by the nonequivalence of several aspects of treatment planning and delivery. A fair comparison should compare similarly advanced delivery techniques from current clinical practice and also assess the robustness of each technique. The present study therefore compared volumetric modulated arc therapy (VMAT) and single-field optimization (SFO) spot scanning proton therapy plans created using a simultaneous integrated boost (SIB) for dose escalation in midesophageal cancer and analyzed the effect of setup and range uncertainties on these plans.

Methods and materials

For 21 patients, SIB plans with a physical dose prescription of 2 Gy or 2.5 Gy/fraction in 25 fractions to planning target volume (PTV)50Gy or PTV62.5Gy (primary tumor with 0.5 cm margins) were created and evaluated for robustness to random setup errors and proton range errors. Dose-volume metrics were compared for the optimal and uncertainty plans, with P<.05 (Wilcoxon) considered significant.

Results

SFO reduced the mean lung dose by 51.4% (range 35.1%-76.1%) and the mean heart dose by 40.9% (range 15.0%-57.4%) compared with VMAT. Proton plan robustness to a 3.5% range error was acceptable. For all patients, the clinical target volume D98 was 95.0% to 100.4% of the prescribed dose and gross tumor volume (GTV) D98 was 98.8% to 101%. Setup error robustness was patient anatomy dependent, and the potential minimum dose per fraction was always lower with SFO than with VMAT. The clinical target volume D98 was lower by 0.6% to 7.8% of the prescribed dose, and the GTV D98 was lower by 0.3% to 2.2% of the prescribed GTV dose.

Conclusions

The SFO plans achieved significant sparing of normal tissue compared with the VMAT plans for midesophageal cancer. The target dose coverage in the SIB proton plans was less robust to random setup errors and might be unacceptable for certain patients. Robust optimization to ensure adequate target coverage of SIB proton plans might be beneficial.

SUBMITTER: Warren S 

PROVIDER: S-EPMC4838670 | biostudies-literature | 2016 May

REPOSITORIES: biostudies-literature

altmetric image

Publications

An Analysis of Plan Robustness for Esophageal Tumors: Comparing Volumetric Modulated Arc Therapy Plans and Spot Scanning Proton Planning.

Warren Samantha S   Partridge Mike M   Bolsi Alessandra A   Lomax Anthony J AJ   Hurt Chris C   Crosby Thomas T   Hawkins Maria A MA  

International journal of radiation oncology, biology, physics 20160130 1


<h4>Purpose</h4>Planning studies to compare x-ray and proton techniques and to select the most suitable technique for each patient have been hampered by the nonequivalence of several aspects of treatment planning and delivery. A fair comparison should compare similarly advanced delivery techniques from current clinical practice and also assess the robustness of each technique. The present study therefore compared volumetric modulated arc therapy (VMAT) and single-field optimization (SFO) spot sc  ...[more]

Similar Datasets

| S-EPMC6612702 | biostudies-literature
| S-EPMC9365973 | biostudies-literature
| S-EPMC8058026 | biostudies-literature
| S-EPMC6560241 | biostudies-literature
| S-EPMC7883303 | biostudies-literature
| S-EPMC5869893 | biostudies-literature
| S-EPMC8853452 | biostudies-literature
| S-EPMC5500391 | biostudies-literature
| S-EPMC5345484 | biostudies-literature
| S-EPMC2967715 | biostudies-other