The Diagnostic Performance of Coronary CT Angiography for the Assessment of Coronary Stenosis in Calcified Plaque.


ABSTRACT:

Purpose

To prospectively evaluate the diagnostic performance of coronary CT angiography (CCTA) for the assessment of coronary stenosis in a calcified plaque, by using conventional coronary angiography (CAG) as a standard reference.

Materials and methods

Eight hundred and ninety-four patients were known to have or have been suspicious of having coronary artery disease, underwent CCTA and conventional coronary angiography (CAG). All the images acquired were assessed. The calcified plaque in CCTA was classified into four types (I-IV) according to the ratio of calcified plaque volume to vessel circumference (RVTC). Overall diagnostic accuracy was made under receiver operating characteristic curve (AUC) analysis. CAG was used as the standard reference.

Results

A total of 12845 segments were evaluated in 894 patients, among which 4955 calcified plaques were detected on 3645(28.4%) segments by CCTA. The overall AUC, sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV) were 0.939, 97.8%, 90.1%, 71.2% and 99.4%, respectively. In type I-II calcification, CCTA had high diagnostic performance in AUC (type I: 0.983; type II: 0.976), sensitivity (96.7%; 98.1%), specificity (99.8%; 97.0%), PPV (95.7%; 90.1%), NPV (99.8%; 99.5%) and accuracy (99.6%; 97.3%). In type III-IV calcification, CCTA has high performance in sensitivity (type III: 97.6%; type IV: 97.9%) and NPV (98.3%; 98.7%), moderate performance in AUC (0.877; 0.829), while remarkable decrease in specificity (78.7%; 67.9%), PPV (71.0%; 56.2%) and accuracy (84.9%; 76.8%).

Conclusion

CCTA has highest accuracy in diagnosing the coronary artery stenosis of type I-II calcified plaques, but has a significant decrease in specificity, PPV and accuracy in type III-IV calcified plaque.

SUBMITTER: Qi L 

PROVIDER: S-EPMC4858195 | biostudies-literature | 2016

REPOSITORIES: biostudies-literature

altmetric image

Publications

The Diagnostic Performance of Coronary CT Angiography for the Assessment of Coronary Stenosis in Calcified Plaque.

Qi Liang L   Tang Li-Jun LJ   Xu Yi Y   Zhu Xiao-Mei XM   Zhang Yu-Dong YD   Shi Hai-Bin HB   Yu Rong-Bin RB  

PloS one 20160505 5


<h4>Purpose</h4>To prospectively evaluate the diagnostic performance of coronary CT angiography (CCTA) for the assessment of coronary stenosis in a calcified plaque, by using conventional coronary angiography (CAG) as a standard reference.<h4>Materials and methods</h4>Eight hundred and ninety-four patients were known to have or have been suspicious of having coronary artery disease, underwent CCTA and conventional coronary angiography (CAG). All the images acquired were assessed. The calcified p  ...[more]

Similar Datasets

| S-EPMC10854174 | biostudies-literature
| S-EPMC4162406 | biostudies-literature
| S-EPMC7977955 | biostudies-literature
| S-EPMC8564016 | biostudies-literature
2022-06-29 | GSE202625 | GEO
| S-EPMC4455680 | biostudies-literature
| S-EPMC4613877 | biostudies-literature
| S-EPMC6328572 | biostudies-literature
| S-EPMC5131836 | biostudies-literature
| S-EPMC10973734 | biostudies-literature