Unknown

Dataset Information

0

Positive Contrast MRI Techniques for Visualization of Iron-Loaded Hernia Mesh Implants in Patients.


ABSTRACT:

Object

In MRI, implants and devices can be delineated via susceptibility artefacts. To discriminate susceptibility voids from proton-free structures, different positive contrast techniques were implemented. The purpose of this study was to evaluate a pulse sequence-based positive contrast technique (PCSI) and a post-processing susceptibility gradient mapping algorithm (SGM) for visualization of iron loaded mesh implants in patients.

Material and methods

Five patients with iron-loaded MR-visible inguinal hernia mesh implants were examined at 1.5 Tesla. A gradient echo sequence (GRE; parameters: TR: 8.3ms; TE: 4.3ms; NSA:2; FA:20°; FOV:350mm²) and a PCSI sequence (parameters: TR: 25ms; TE: 4.6ms; NSA:4; FA:20°; FOV:350mm²) with on-resonant proton suppression were performed. SGM maps were calculated using two algorithms. Image quality and mesh delineation were independently evaluated by three radiologists.

Results

On GRE, the iron-loaded meshes generated distinct susceptibility-induced signal voids. PCSI exhibited susceptibility differences including the meshes as hyperintense signals. SGM exhibited susceptibility differences with positive contrast. Visually, the different algorithms presented no significant differences. Overall, the diagnostic value was rated best in GRE whereas PCSI and SGM were barely "sufficient".

Conclusion

Both "positive contrast" techniques depicted implanted meshes with hyperintense signal. SGM comes without additional acquisition time and can therefore be utilized in every patient.

SUBMITTER: Ciritsis A 

PROVIDER: S-EPMC4871409 | biostudies-literature | 2016

REPOSITORIES: biostudies-literature

altmetric image

Publications

Positive Contrast MRI Techniques for Visualization of Iron-Loaded Hernia Mesh Implants in Patients.

Ciritsis Alexander A   Truhn Daniel D   Hansen Nienke L NL   Otto Jens J   Kuhl Christiane K CK   Kraemer Nils A NA  

PloS one 20160518 5


<h4>Object</h4>In MRI, implants and devices can be delineated via susceptibility artefacts. To discriminate susceptibility voids from proton-free structures, different positive contrast techniques were implemented. The purpose of this study was to evaluate a pulse sequence-based positive contrast technique (PCSI) and a post-processing susceptibility gradient mapping algorithm (SGM) for visualization of iron loaded mesh implants in patients.<h4>Material and methods</h4>Five patients with iron-loa  ...[more]

Similar Datasets

| S-EPMC4967714 | biostudies-literature
| S-EPMC5338531 | biostudies-literature
| S-EPMC6124042 | biostudies-literature
| S-EPMC3170541 | biostudies-literature
| S-EPMC5674517 | biostudies-literature
| S-EPMC6510773 | biostudies-literature
| S-EPMC3429787 | biostudies-literature