Fundamental Importance of Reference Glucose Analyzer Accuracy for Evaluating the Performance of Blood Glucose Monitoring Systems (BGMSs).


ABSTRACT:

Background

As blood glucose monitoring system (BGMS) accuracy is based on comparison of BGMS and laboratory reference glucose analyzer results, reference instrument accuracy is important to discriminate small differences between BGMS and reference glucose analyzer results. Here, we demonstrate the important role of reference glucose analyzer accuracy in BGMS accuracy evaluations.

Methods

Two clinical studies assessed the performance of a new BGMS, using different reference instrument procedures. BGMS and YSI analyzer results were compared for fingertip blood that was obtained by untrained subjects' self-testing and study staff testing, respectively. YSI analyzer accuracy was monitored using traceable serum controls.

Results

In study 1 (N = 136), 94.1% of BGMS results were within International Organization for Standardization (ISO) 15197:2013 accuracy criteria; YSI analyzer serum control results showed a negative bias (-0.64% to -2.48%) at the first site and a positive bias (3.36% to 6.91%) at the other site. In study 2 (N = 329), 97.8% of BGMS results were within accuracy criteria; serum controls showed minimal bias (<0.92%) at both sites.

Conclusions

These findings suggest that the ability to demonstrate that a BGMS meets accuracy guidelines is influenced by reference instrument accuracy.

SUBMITTER: Bailey TS 

PROVIDER: S-EPMC4928230 | biostudies-literature | 2016 Jul

REPOSITORIES: biostudies-literature

altmetric image

Publications

Fundamental Importance of Reference Glucose Analyzer Accuracy for Evaluating the Performance of Blood Glucose Monitoring Systems (BGMSs).

Bailey Timothy S TS   Klaff Leslie J LJ   Wallace Jane F JF   Greene Carmine C   Pardo Scott S   Harrison Bern B   Simmons David A DA  

Journal of diabetes science and technology 20160628 4


<h4>Background</h4>As blood glucose monitoring system (BGMS) accuracy is based on comparison of BGMS and laboratory reference glucose analyzer results, reference instrument accuracy is important to discriminate small differences between BGMS and reference glucose analyzer results. Here, we demonstrate the important role of reference glucose analyzer accuracy in BGMS accuracy evaluations.<h4>Methods</h4>Two clinical studies assessed the performance of a new BGMS, using different reference instrum  ...[more]

Similar Datasets

| S-EPMC3558677 | biostudies-literature
| S-EPMC3696921 | biostudies-literature
| S-EPMC9445332 | biostudies-literature
| S-EPMC6080122 | biostudies-literature
| S-EPMC4575307 | biostudies-literature
| S-EPMC6104257 | biostudies-literature
| S-EPMC6203032 | biostudies-literature
| S-EPMC5950999 | biostudies-literature
| S-EPMC4609557 | biostudies-literature
| S-EPMC10838869 | biostudies-literature