Adverse reaction to metal debris in a patient with acetabular shell loosening 8 years after ceramic-on-metal total hip arthroplasty.


ABSTRACT: A 41-year-old woman presented 8 years after a left total hip arthroplasty. She complained of progressive groin pain for several months. Radiographs demonstrated a hard-on-hard bearing surface combination and radiolucent lines surrounding the acetabular shell. Laboratory analysis revealed a mild leukocytosis, a normal erythrocyte sedimentation rate, and a mildly elevated C-reactive protein. Serum cobalt and chromium levels were markedly elevated. Aspiration of the hip joint was negative for infection. Magnetic resonance imaging failed to demonstrate a pseudotumor. Revision total hip arthroplasty was performed, and a ceramic-on-metal bearing surface combination was explanted. Significant intraoperative findings included dark gray synovial fluid, metal transfer onto the ceramic femoral head, and a grossly loose acetabular shell pivoting about a single well-fixed screw. The explanted components otherwise appeared normal macroscopically. Histologic analysis of the capsular tissue demonstrated aseptic lymphocyte-dominated vasculitis-associated lesion and inclusion bodies consistent with third-body wear. Revision arthroplasty to a ceramic-on-polyethylene bearing surface combination was performed with a good clinical result and laboratory normalization at 9-month follow-up.

SUBMITTER: Pulley BR 

PROVIDER: S-EPMC4958114 | biostudies-literature | 2015 Dec

REPOSITORIES: biostudies-literature

altmetric image

Publications

Adverse reaction to metal debris in a patient with acetabular shell loosening 8 years after ceramic-on-metal total hip arthroplasty.

Pulley Benjamin R BR   Trinh Thai Q TQ   Bentley Jared C JC   Politi Joel R JR  

Arthroplasty today 20151021 4


A 41-year-old woman presented 8 years after a left total hip arthroplasty. She complained of progressive groin pain for several months. Radiographs demonstrated a hard-on-hard bearing surface combination and radiolucent lines surrounding the acetabular shell. Laboratory analysis revealed a mild leukocytosis, a normal erythrocyte sedimentation rate, and a mildly elevated C-reactive protein. Serum cobalt and chromium levels were markedly elevated. Aspiration of the hip joint was negative for infec  ...[more]

Similar Datasets

| S-EPMC5712035 | biostudies-literature
| S-EPMC4958112 | biostudies-literature
| S-EPMC7797157 | biostudies-literature
| S-EPMC5485220 | biostudies-literature
| S-EPMC9874008 | biostudies-literature
| S-EPMC5365402 | biostudies-literature
| S-EPMC7502561 | biostudies-literature
| S-EPMC2806984 | biostudies-literature
| S-EPMC10517284 | biostudies-literature
| S-EPMC9157011 | biostudies-literature