Unknown

Dataset Information

0

Clinical Evaluation of the BD FACSPresto™ Near-Patient CD4 Counter in Kenya.


ABSTRACT: The BD FACSPresto™ Near-Patient CD4 Counter was developed to expand HIV/AIDS management in resource-limited settings. It measures absolute CD4 counts (AbsCD4), percent CD4 (%CD4), and hemoglobin (Hb) from a single drop of capillary or venous blood in approximately 23 minutes, with throughput of 10 samples per hour. We assessed the performance of the BD FACSPresto system, evaluating accuracy, stability, linearity, precision, and reference intervals using capillary and venous blood at KEMRI/CDC HIV-research laboratory, Kisumu, Kenya, and precision and linearity at BD Biosciences, California, USA.For accuracy, venous samples were tested using the BD FACSCalibur™ instrument with BD Tritest™ CD3/CD4/CD45 reagent, BD Trucount™ tubes, and BD Multiset™ software for AbsCD4 and %CD4, and the Sysmex™ KX-21N for Hb. Stability studies evaluated duration of staining (18-120-minute incubation), and effects of venous blood storage <6-24 hours post-draw. A normal cohort was tested for reference intervals. Precision covered multiple days, operators, and instruments. Linearity required mixing two pools of samples, to obtain evenly spaced concentrations for AbsCD4, total lymphocytes, and Hb.AbsCD4 and %CD4 venous/capillary (N = 189/ N = 162) accuracy results gave Deming regression slopes within 0.97-1.03 and R2 ?0.96. For Hb, Deming regression results were R2 ?0.94 and slope ?0.94 for both venous and capillary samples. Stability varied within 10% 2 hours after staining and for venous blood stored less than 24 hours. Reference intervals results showed that gender-but not age-differences were statistically significant (p<0.05). Precision results had <3.5% coefficient of variation for AbsCD4, %CD4, and Hb, except for low AbsCD4 samples (<6.8%). Linearity was 42-4,897 cells/?L for AbsCD4, 182-11,704 cells/?L for total lymphocytes, and 2-24 g/dL for Hb.The BD FACSPresto system provides accurate, precise clinical results for capillary or venous blood samples and is suitable for near-patient CD4 testing.ClinicalTrials.gov NCT02396355.

SUBMITTER: Angira F 

PROVIDER: S-EPMC4970792 | biostudies-literature | 2016

REPOSITORIES: biostudies-literature

altmetric image

Publications

Clinical Evaluation of the BD FACSPresto™ Near-Patient CD4 Counter in Kenya.

Angira Francis F   Akoth Benta B   Omolo Paul P   Opollo Valarie V   Bornheimer Scott S   Judge Kevin K   Tilahun Henok H   Lu Beverly B   Omana-Zapata Imelda I   Zeh Clement C  

PloS one 20160802 8


<h4>Background</h4>The BD FACSPresto™ Near-Patient CD4 Counter was developed to expand HIV/AIDS management in resource-limited settings. It measures absolute CD4 counts (AbsCD4), percent CD4 (%CD4), and hemoglobin (Hb) from a single drop of capillary or venous blood in approximately 23 minutes, with throughput of 10 samples per hour. We assessed the performance of the BD FACSPresto system, evaluating accuracy, stability, linearity, precision, and reference intervals using capillary and venous bl  ...[more]

Similar Datasets

| S-EPMC5444776 | biostudies-literature
| S-EPMC5407647 | biostudies-literature
| S-EPMC5730955 | biostudies-literature
| S-EPMC4880207 | biostudies-literature
| S-EPMC3763498 | biostudies-literature
| S-EPMC3725734 | biostudies-literature
| S-EPMC8376527 | biostudies-literature
| S-EPMC7658244 | biostudies-literature
| S-EPMC5886872 | biostudies-literature
| S-EPMC6899525 | biostudies-literature