Comparison of pre-processing methods for multiplex bead-based immunoassays.


ABSTRACT:

Background

High throughput protein expression studies can be performed using bead-based protein immunoassays, such as the Luminex® xMAP® technology. Technical variability is inherent to these experiments and may lead to systematic bias and reduced power. To reduce technical variability, data pre-processing is performed. However, no recommendations exist for the pre-processing of Luminex® xMAP® data.

Results

We compared 37 different data pre-processing combinations of transformation and normalization methods in 42 samples on 384 analytes obtained from a multiplex immunoassay based on the Luminex® xMAP® technology. We evaluated the performance of each pre-processing approach with 6 different performance criteria. Three performance criteria were plots. All plots were evaluated by 15 independent and blinded readers. Four different combinations of transformation and normalization methods performed well as pre-processing procedure for this bead-based protein immunoassay.

Conclusions

The following combinations of transformation and normalization were suitable for pre-processing Luminex® xMAP® data in this study: weighted Box-Cox followed by quantile or robust spline normalization (rsn), asinh transformation followed by loess normalization and Box-Cox followed by rsn.

SUBMITTER: Rausch TK 

PROVIDER: S-EPMC4982217 | biostudies-literature | 2016 Aug

REPOSITORIES: biostudies-literature

altmetric image

Publications

Comparison of pre-processing methods for multiplex bead-based immunoassays.

Rausch Tanja K TK   Schillert Arne A   Ziegler Andreas A   Lüking Angelika A   Zucht Hans-Dieter HD   Schulz-Knappe Peter P  

BMC genomics 20160811 1


<h4>Background</h4>High throughput protein expression studies can be performed using bead-based protein immunoassays, such as the Luminex® xMAP® technology. Technical variability is inherent to these experiments and may lead to systematic bias and reduced power. To reduce technical variability, data pre-processing is performed. However, no recommendations exist for the pre-processing of Luminex® xMAP® data.<h4>Results</h4>We compared 37 different data pre-processing combinations of transformatio  ...[more]

Similar Datasets

| S-EPMC7546899 | biostudies-literature
| S-EPMC5736219 | biostudies-literature
| S-EPMC5685631 | biostudies-literature
| S-EPMC2689848 | biostudies-other
| S-EPMC1392679 | biostudies-literature
| S-EPMC3187343 | biostudies-literature
| S-EPMC9783081 | biostudies-literature
| S-EPMC4762227 | biostudies-literature
| S-EPMC2045348 | biostudies-literature
| S-EPMC10654656 | biostudies-literature