Protein Quantification by Derivatization-Free High-Performance Liquid Chromatography of Aromatic Amino Acids.


ABSTRACT: Amino acid analysis is considered to be the gold standard for quantitative peptide and protein analysis. Here, we would like to propose a simple HPLC/UV method based on a reversed-phase separation of the aromatic amino acids tyrosine (Tyr), phenylalanine (Phe), and optionally tryptophan (Trp) without any derivatization. The hydrolysis of the proteins and peptides was performed by an accelerated microwave technique, which needs only 30 minutes. Two internal standard compounds, homotyrosine (HTyr) and 4-fluorophenylalanine (FPhe) were used for calibration. The limit of detection (LOD) was estimated to be 0.05 µM (~10?µg/L) for tyrosine and phenylalanine at 215?nm. The LOD for a protein determination was calculated to be below 16?mg/L (~300?ng BSA absolute). Aromatic amino acid analysis (AAAA) offers excellent accuracy and a precision of about 5% relative standard deviation, including the hydrolysis step. The method was validated with certified reference materials (CRM) of amino acids and of a pure protein (bovine serum albumin, BSA). AAAA can be used for the quantification of aromatic amino acids, isolated peptides or proteins, complex peptide or protein samples, such as serum or milk powder, and peptides or proteins immobilized on solid supports.

SUBMITTER: Hesse A 

PROVIDER: S-EPMC4983383 | biostudies-literature | 2016

REPOSITORIES: biostudies-literature

altmetric image

Publications

Protein Quantification by Derivatization-Free High-Performance Liquid Chromatography of Aromatic Amino Acids.

Hesse Almut A   Weller Michael G MG  

Journal of amino acids 20160731


Amino acid analysis is considered to be the gold standard for quantitative peptide and protein analysis. Here, we would like to propose a simple HPLC/UV method based on a reversed-phase separation of the aromatic amino acids tyrosine (Tyr), phenylalanine (Phe), and optionally tryptophan (Trp) without any derivatization. The hydrolysis of the proteins and peptides was performed by an accelerated microwave technique, which needs only 30 minutes. Two internal standard compounds, homotyrosine (HTyr)  ...[more]

Similar Datasets

| S-EPMC9135559 | biostudies-literature
| S-EPMC6648825 | biostudies-literature
| S-EPMC3976834 | biostudies-literature
| S-EPMC6273726 | biostudies-literature
| S-EPMC2848510 | biostudies-literature
| S-EPMC5431165 | biostudies-literature
| S-EPMC8697017 | biostudies-literature
| S-EPMC7045947 | biostudies-literature
| S-EPMC5605124 | biostudies-literature
| S-EPMC6854794 | biostudies-literature