Ontology highlight
ABSTRACT: Introduction
We developed a device (FreeO2) that automatically adjusts the oxygen flow rates based on patients' needs, in order to limit hyperoxia and hypoxemia and to automatically wean them from oxygen.Objective
The aim of this study was to evaluate the feasibility of using FreeO2 in patients hospitalized in the respiratory ward for an acute exacerbation of COPD.Methods
We conducted a randomized controlled trial comparing FreeO2 vs manual oxygen titration in the respiratory ward of a university hospital. We measured the perception of appropriateness of oxygen titration and monitoring in both groups by nurses and attending physicians using a Likert scale. We evaluated the time in the target range of oxygen saturation (SpO2) as defined for each patient by the attending physician, the time with severe desaturation (SpO2 <85%), and the time with hyperoxia (SpO2 >5% above the target). We also recorded length of stay, intensive care unit admissions, and readmission rate. Fifty patients were randomized (25 patients in both groups; mean age: 72±8 years; mean forced expiratory volume in 1 second: 1.00±0.49 L; and mean initial O2 flow 2.0±1.0 L/min).Results
Nurses and attending physicians felt that oxygen titration and monitoring were equally appropriate with both O2 administration systems. The percentage of time within the SpO2 target was significantly higher with FreeO2, and the time with severe desaturation and hyperoxia was significantly reduced with FreeO2. Time from study inclusion to hospital discharge was 5.8±4.4 days with FreeO2 and 8.4±6.0 days with usual oxygen administration (P=0.051).Conclusion
FreeO2 was deemed as an appropriate oxygen administration system by nurses and physicians of a respiratory unit. This system maintained SpO2 at the target level better than did manual titration and reduced periods of desaturation and hyperoxia. Our results also suggest that FreeO2 has the potential to reduce the hospital length of stay.
SUBMITTER: Lellouche F
PROVIDER: S-EPMC5003517 | biostudies-literature | 2016
REPOSITORIES: biostudies-literature
Lellouche François F Bouchard Pierre-Alexandre PA Roberge Maude M Simard Serge S L'Her Erwan E Maltais François F Lacasse Yves Y
International journal of chronic obstructive pulmonary disease 20160824
<h4>Introduction</h4>We developed a device (FreeO2) that automatically adjusts the oxygen flow rates based on patients' needs, in order to limit hyperoxia and hypoxemia and to automatically wean them from oxygen.<h4>Objective</h4>The aim of this study was to evaluate the feasibility of using FreeO2 in patients hospitalized in the respiratory ward for an acute exacerbation of COPD.<h4>Methods</h4>We conducted a randomized controlled trial comparing FreeO2 vs manual oxygen titration in the respira ...[more]