Sahoo Malaya K MK Varghese Vici V White Elizabeth E Winslow Meg M Katzenstein David A DA Shafer Robert W RW Pinsky Benjamin A BA
Journal of clinical microbiology 20160817 10
HIV-1 RNA quantitation in plasma, or virus load testing, is the primary method by which the response to antiretroviral therapy is monitored. Here we describe evaluation of the Aptima HIV-1 Quant Dx assay (Aptima) performed on the automated Panther system. The clinical performance of Aptima was compared to that of the Cobas AmpliPrep/Cobas TaqMan HIV-1 Test v2.0 (CAP/CTM) using 162 EDTA plasma samples collected from patients undergoing HIV-1 monitoring. Overall agreement was 84.0% (136/162), with ...[more]