The performance of single and multi-collector ICP-MS instruments for fast and reliable 34S/32S isotope ratio measurements.


ABSTRACT: The performance and validation characteristics of different single collector inductively coupled plasma mass spectrometers based on different technical principles (ICP-SFMS, ICP-QMS in reaction and collision modes, and ICP-MS/MS) were evaluated in comparison to the performance of MC ICP-MS for fast and reliable S isotope ratio measurements. The validation included the determination of LOD, BEC, measurement repeatability, within-lab reproducibility and deviation from certified values as well as a study on instrumental isotopic fractionation (IIF) and the calculation of the combined standard measurement uncertainty. Different approaches of correction for IIF applying external intra-elemental IIF correction (aka standard-sample bracketing) using certified S reference materials and internal inter-elemental IIF (aka internal standardization) correction using Si isotope ratios in MC ICP-MS are explained and compared. The resulting combined standard uncertainties of examined ICP-QMS systems were not better than 0.3-0.5% (uc,rel), which is in general insufficient to differentiate natural S isotope variations. Although the performance of the single collector ICP-SFMS is better (single measurement uc,rel = 0.08%), the measurement reproducibility (>0.2%) is the major limit of this system and leaves room for improvement. MC ICP-MS operated in the edge mass resolution mode, applying bracketing for correction of IIF, provided isotope ratio values with the highest quality (relative combined measurement uncertainty: 0.02%; deviation from the certified value: <0.002%).

SUBMITTER: Hanousek O 

PROVIDER: S-EPMC5087850 | biostudies-literature | 2016 Nov

REPOSITORIES: biostudies-literature

altmetric image

Publications

The performance of single and multi-collector ICP-MS instruments for fast and reliable <sup>34</sup>S/<sup>32</sup>S isotope ratio measurements.

Hanousek Ondrej O   Brunner Marion M   Pröfrock Daniel D   Irrgeher Johanna J   Prohaska Thomas T  

Analytical methods : advancing methods and applications 20160923 42


The performance and validation characteristics of different single collector inductively coupled plasma mass spectrometers based on different technical principles (ICP-SFMS, ICP-QMS in reaction and collision modes, and ICP-MS/MS) were evaluated in comparison to the performance of MC ICP-MS for fast and reliable S isotope ratio measurements. The validation included the determination of LOD, BEC, measurement repeatability, within-lab reproducibility and deviation from certified values as well as a  ...[more]

Similar Datasets

| S-EPMC5366171 | biostudies-literature
| S-EPMC9314310 | biostudies-literature
| S-EPMC3589628 | biostudies-literature
| S-EPMC10909487 | biostudies-literature
| S-EPMC8041249 | biostudies-literature
| S-EPMC9982939 | biostudies-literature
| S-EPMC8405517 | biostudies-literature
| S-EPMC8783016 | biostudies-literature
| S-EPMC7200944 | biostudies-literature
| S-EPMC9132797 | biostudies-literature