Efficacy of an Overnight Predictive Low-Glucose Suspend System in Relation to Hypoglycemia Risk Factors in Youth and Adults With Type 1 Diabetes.


ABSTRACT: BACKGROUND:We developed a system to suspend insulin pump delivery overnight when the glucose trend predicts hypoglycemia. This predictive low-glucose suspend (PLGS) system substantially reduces nocturnal hypoglycemia without an increase in morning ketosis. Evaluation of hypoglycemia risk factors that could potentially influence the efficacy of the system remains critical for understanding possible problems with the system and identifying patients that may have the greatest benefit when using the system. METHODS:The at-home randomized trial consisted of 127 study participants with hemoglobin A1c (A1C) of ?8.5% (mmol/mol) for patients aged 4-14 years and ?8.0% for patient aged 15-45 years. Factors assessed included age, gender, A1C, diabetes duration, daily percentage basal insulin, total daily dose of insulin (units/kg-day), bedtime BG, bedtime snack, insulin on board, continuous glucose monitor (CGM) rate of change (ROC), day of the week, time system activated, daytime exercise intensity, and daytime CGM-measured hypoglycemia. RESULTS:The PLGS system was effective in preventing hypoglycemia for each factor subgroup. There was no evidence that the PLGS system was more or less effective in preventing hypoglycemia in any one subgroup compared with the other subgroups based on that factor. In addition, the effect of the system on overnight hyperglycemia did not differ in subgroups. CONCLUSIONS:The PLGS system tested in this study effectively reduced hypoglycemia without a meaningful increase in hyperglycemia across a variety of factors.

SUBMITTER: Calhoun PM 

PROVIDER: S-EPMC5094319 | biostudies-literature | 2016 Nov

REPOSITORIES: biostudies-literature

altmetric image

Publications

Efficacy of an Overnight Predictive Low-Glucose Suspend System in Relation to Hypoglycemia Risk Factors in Youth and Adults With Type 1 Diabetes.

Calhoun Peter M PM   Buckingham Bruce A BA   Maahs David M DM   Hramiak Irene I   Wilson Darrell M DM   Aye Tandy T   Clinton Paula P   Chase Peter P   Messer Laurel L   Kollman Craig C   Beck Roy W RW   Lum John J  

Journal of diabetes science and technology 20161101 6


<h4>Background</h4>We developed a system to suspend insulin pump delivery overnight when the glucose trend predicts hypoglycemia. This predictive low-glucose suspend (PLGS) system substantially reduces nocturnal hypoglycemia without an increase in morning ketosis. Evaluation of hypoglycemia risk factors that could potentially influence the efficacy of the system remains critical for understanding possible problems with the system and identifying patients that may have the greatest benefit when u  ...[more]

Similar Datasets

| S-EPMC5319476 | biostudies-literature
| S-EPMC8252907 | biostudies-literature
| S-EPMC5086306 | biostudies-literature
| S-EPMC3994933 | biostudies-literature
| S-EPMC3746249 | biostudies-literature
| S-EPMC6708266 | biostudies-literature
| S-EPMC8655276 | biostudies-literature
| S-EPMC2768224 | biostudies-literature
| S-EPMC3041185 | biostudies-literature
| S-EPMC6208158 | biostudies-literature