Unknown

Dataset Information

0

Corrosion Damage and Wear Mechanisms in Long-Term Retrieved CoCr Femoral Components for Total Knee Arthroplasty.


ABSTRACT: Metal debris and ion release has raised concerns in joint arthroplasty. The purpose of this study was to characterize the sources of metallic ions and particulate debris released from long-term (in vivo >15 years) total knee arthroplasty femoral components.A total of 52 CoCr femoral condyles were identified as having been implanted for more than 15 years. The femoral components were examined for incidence of 5 types of damage (metal-on-metal wear due to historical polyethylene insert failure, mechanically assisted crevice corrosion at taper interfaces, cement interface corrosion, third-body abrasive wear, and inflammatory cell-induced corrosion [ICIC]). Third-body abrasive wear was evaluated using the Hood method for polyethylene components and a similar method quantifying surface damage of the femoral condyle was used. The total area damaged by ICIC was quantified using digital photogrammetry.Surface damage associated with corrosion and/or CoCr debris release was identified in 51 (98%) CoCr femoral components. Five types of damage were identified: 98% of femoral components exhibited third-body abrasive wear (mostly observed as scratching, n = 51/52), 29% of femoral components exhibited ICIC damage (n = 15/52), 41% exhibited cement interface damage (n = 11/27), 17% exhibited metal-on-metal wear after wear-through of the polyethylene insert (n = 9/52), and 50% of the modular femoral components exhibited mechanically assisted crevice corrosion taper damage (n = 2/4). The total ICIC-damaged area was an average of 0.11 ± 0.12 mm2 (range: 0.01-0.46 mm2).Although implant damage in total knee arthroplasty is typically reported with regard to the polyethylene insert, the results of this study demonstrate that abrasive and corrosive damage occurs on the CoCr femoral condyle in vivo.

SUBMITTER: Arnholt CM 

PROVIDER: S-EPMC5107165 | biostudies-literature | 2016 Dec

REPOSITORIES: biostudies-literature

altmetric image

Publications

Corrosion Damage and Wear Mechanisms in Long-Term Retrieved CoCr Femoral Components for Total Knee Arthroplasty.

Arnholt Christina M CM   MacDonald Daniel W DW   Malkani Arthur L AL   Klein Gregg R GR   Rimnac Clare M CM   Kurtz Steven M SM   Kocagoz Sevi B SB   Gilbert Jeremy L JL  

The Journal of arthroplasty 20160512 12


<h4>Background</h4>Metal debris and ion release has raised concerns in joint arthroplasty. The purpose of this study was to characterize the sources of metallic ions and particulate debris released from long-term (in vivo >15 years) total knee arthroplasty femoral components.<h4>Methods</h4>A total of 52 CoCr femoral condyles were identified as having been implanted for more than 15 years. The femoral components were examined for incidence of 5 types of damage (metal-on-metal wear due to histori  ...[more]

Similar Datasets

| S-EPMC9085432 | biostudies-literature
| S-EPMC4956686 | biostudies-literature
2010-04-12 | E-GEOD-21164 | biostudies-arrayexpress
| S-EPMC11303532 | biostudies-literature
| S-EPMC5298367 | biostudies-literature
| S-EPMC3971237 | biostudies-literature
| S-EPMC5247518 | biostudies-literature
2010-04-02 | GSE21164 | GEO
| S-EPMC8079325 | biostudies-literature
| S-EPMC4956687 | biostudies-literature