ABSTRACT: Aims
Managing bladder pressure in patients with neurogenic bladders is needed to improve rehabilitation options, avoid upper tract damage, incontinence, and their associated co-morbidities and mortality. Current methods of determining bladder contractions are not amenable to chronic or ambulatory settings. In this study we evaluated detection of bladder contractions using a novel piezoelectric catheter-free pressure sensor placed in a suburothelial bladder location in animals.Methods
Wired prototypes of the pressure monitor were implanted into 2 nonsurvival (feline and canine) and one 13-day survival (canine) animal. Vesical pressures were obtained from the device in both suburothelial and intraluminal locations and simultaneously from a pressure sensing catheter in the bladder. Intravesical pressure was monitored in the survival animal over 10 days from the suburothelial location and necropsy was performed to assess migration and erosion.Results
In the nonsurvival animals, the average correlation between device and reference catheter data was high during both electrically stimulated bladder contractions and manual compressions (r = 0.93±0.03, r = 0.89±0.03). Measured pressures correlated strongly (r = 0.98±0.02) when the device was placed in the bladder lumen. The survival animal initially recorded physiologic data, but later this deteriorated. However, endstage intraluminal device recordings correlated (r = 0.85±0.13) with the pressure catheter. Significant erosion of the implant through the detrusor was found.Conclusions
This study confirms correlation between suburothelial pressure readings and intravesical bladder pressures. Due to device erosion during ambulatory studies, a wireless implant is recommended for clinical rehabilitation applications.
SUBMITTER: Majerus SJ
PROVIDER: S-EPMC5218553 | biostudies-literature | 2017
REPOSITORIES: biostudies-literature
Majerus Steve J A SJ Fletter Paul C PC Ferry Elizabeth K EK Zhu Hui H Gustafson Kenneth J KJ Damaser Margot S MS
PloS one 20170106 1
<h4>Aims</h4>Managing bladder pressure in patients with neurogenic bladders is needed to improve rehabilitation options, avoid upper tract damage, incontinence, and their associated co-morbidities and mortality. Current methods of determining bladder contractions are not amenable to chronic or ambulatory settings. In this study we evaluated detection of bladder contractions using a novel piezoelectric catheter-free pressure sensor placed in a suburothelial bladder location in animals.<h4>Methods ...[more]