Unknown

Dataset Information

0

Interlaboratory Variability in Plasma Creatinine Measurement and the Relation with Estimated Glomerular Filtration Rate and Chronic Kidney Disease Diagnosis.


ABSTRACT:

Background and objectives

The tracing of creatinine (Cr) reference materials to isotope dilution mass spectrometry-assigned values was implemented worldwide to reduce interlaboratory variability and improve assay accuracy. The aims of this study were to examine the current extent of interlaboratory variability and its effect on eGFR.

Design, setting, participants, & measurements

Leftover plasma from 2-3 consecutive days was obtained from 53 intensive care unit patients with a range of kidney functions. Individual patient samples were pooled and split and sent to 12 different laboratories for Cr measurement. For each patient, the mean Cr and Chronic Kidney Disease Epidemiology Collaboration eGFR (eGFR-EPI), assuming a 65-year-old nonblack woman, were determined. Interlaboratory variability was assessed by the range and SD of Cr and eGFR-EPI. This was repeated after stratifying by assay type and by the median Cr of 1.36 mg/dl. For patients whose eGFR-EPI range included 60 ml/min per 1.73 m2, the percentage of laboratories with eGFR-EPI<60 ml/min per 1.73 m2 was determined.

Results

The mean±SD of the Cr and eGFR-EPI ranges were 0.20±0.09 mg/dl and 14±9 ml/min per 1.73 m2 for Cr<1.36 mg/dl. Jaffe Cr results were an average 0.1 mg/dl (Cr?1.36 mg/dl) and 0.05 mg/dl (Cr<1.36 mg/dl) higher than enzymatic results (P<0.001 for both). Ten patients had an eGFR-EPI range that included 60 ml/min per 1.73 m2. Their median eGFR-EPI range was 15 ml/min per 1.73 m2. There was significant discordance in the diagnosis of CKD (eGFR-EPI<60 ml/min per 1.73 m2), with laboratories using Jaffe Cr methods making the diagnosis more frequently than those using enzymatic Cr methods (60% versus 39%).

Conclusions

Significant interlaboratory variability in Cr measurement still exists. Jaffe assays yield higher Cr values than enzymatic assays, leading to lower eGFR-EPIs and more frequent CKD diagnoses. Further improvements in assay performance are required to standardize patient CKD diagnosis and to facilitate longitudinal Cr monitoring across laboratories.

SUBMITTER: Lee E 

PROVIDER: S-EPMC5220660 | biostudies-literature | 2017 Jan

REPOSITORIES: biostudies-literature

altmetric image

Publications

Interlaboratory Variability in Plasma Creatinine Measurement and the Relation with Estimated Glomerular Filtration Rate and Chronic Kidney Disease Diagnosis.

Lee Elizabeth E   Collier Christine P CP   White Christine A CA  

Clinical journal of the American Society of Nephrology : CJASN 20161108 1


<h4>Background and objectives</h4>The tracing of creatinine (Cr) reference materials to isotope dilution mass spectrometry-assigned values was implemented worldwide to reduce interlaboratory variability and improve assay accuracy. The aims of this study were to examine the current extent of interlaboratory variability and its effect on eGFR.<h4>Design, setting, participants, & measurements</h4>Leftover plasma from 2-3 consecutive days was obtained from 53 intensive care unit patients with a rang  ...[more]

Similar Datasets

| 2300288 | ecrin-mdr-crc
| S-EPMC5518963 | biostudies-other
| S-EPMC4726452 | biostudies-literature
| S-EPMC6334920 | biostudies-literature
| S-EPMC9687311 | biostudies-literature
| S-EPMC9258722 | biostudies-literature
| S-EPMC4689117 | biostudies-literature
| S-EPMC5796440 | biostudies-literature
| S-EPMC7439334 | biostudies-literature
| S-EPMC10545821 | biostudies-literature