A comparison of the Airtraq®, McGrath®, and Macintosh laryngoscopes for difficult paediatric intubation: A manikin study.


ABSTRACT:

Background

The efficacy of devices for difficult intubation in paediatric patients, especially with a Cormack-Lehane grade 4 view, has yet to be established. We compared intubating parameters among three devices (the Airtraq®, McGrath®, and Macintosh laryngoscopes).

Methods

This study is a randomised cross-over trial. Participants were 20 anaesthetists. Each device was tested three times using a paediatric manikin with a Cormack-Lehane grade 4 view. The order to use each device was randomised by a computer-generated random sequence. The primary endpoint was the rate of successful intubation. Secondary endpoints included the time taken to intubate, percentage of glottic opening score, and severity of potential dental trauma.

Results

The successful intubation rates of the Airtraq®, McGrath®, and Macintosh laryngoscopes were 100%, 72%, and 45%, respectively. The risk ratio of the success rates of Airtraq® compared with McGrath® and Macintosh laryngoscopes were 1.40 (95% CI; 1.19-1.64, P < 0.001) and 2.22 (95% CI; 1.68-2.94, P < 0.001), respectively. The modified Cormack-Lehane grade and percentage of the glottic opening score were better for the Airtraq® than for the other devices. The dental trauma score was lower for the Airtraq® than for the other devices. There were no significant differences in the intubation time among the groups.

Conclusions

The Airtraq® had higher success rate, had better visibility, and was associated with less dental trauma than the other devices in a difficult paediatric intubation model with a Cormack-Lehane grade 4 view.

SUBMITTER: Owada G 

PROVIDER: S-EPMC5302788 | biostudies-literature | 2017

REPOSITORIES: biostudies-literature

altmetric image

Publications

A comparison of the Airtraq®, McGrath®, and Macintosh laryngoscopes for difficult paediatric intubation: A manikin study.

Owada Gen G   Mihara Takahiro T   Inagawa Gaku G   Asakura Ayako A   Goto Takahisa T   Ka Koui K  

PloS one 20170210 2


<h4>Background</h4>The efficacy of devices for difficult intubation in paediatric patients, especially with a Cormack-Lehane grade 4 view, has yet to be established. We compared intubating parameters among three devices (the Airtraq®, McGrath®, and Macintosh laryngoscopes).<h4>Methods</h4>This study is a randomised cross-over trial. Participants were 20 anaesthetists. Each device was tested three times using a paediatric manikin with a Cormack-Lehane grade 4 view. The order to use each device wa  ...[more]

Similar Datasets

| S-EPMC5040779 | biostudies-literature
| S-EPMC4110170 | biostudies-literature
| S-EPMC6756366 | biostudies-literature
| S-EPMC6255773 | biostudies-literature
| S-EPMC8731221 | biostudies-literature
| S-EPMC8234758 | biostudies-literature
| S-EPMC4623092 | biostudies-literature
| S-EPMC5520596 | biostudies-other
| S-EPMC7441605 | biostudies-literature
| S-EPMC6717380 | biostudies-literature