The Effects of Breathing Motion on DCE-MRI Images: Phantom Studies Simulating Respiratory Motion to Compare CAIPIRINHA-VIBE, Radial-VIBE, and Conventional VIBE.


ABSTRACT:

Objective

To compare the breathing effects on dynamic contrast-enhanced (DCE)-MRI between controlled aliasing in parallel imaging results in higher acceleration (CAIPIRINHA)-volumetric interpolated breath-hold examination (VIBE), radial VIBE with k-space-weighted image contrast view-sharing (radial-VIBE), and conventional VIBE (c-VIBE) sequences using a dedicated phantom experiment.

Materials and methods

We developed a moving platform to simulate breathing motion. We conducted dynamic scanning on a 3T machine (MAGNETOM Skyra, Siemens Healthcare) using CAIPIRINHA-VIBE, radial-VIBE, and c-VIBE for six minutes per sequence. We acquired MRI images of the phantom in both static and moving modes, and we also obtained motion-corrected images for the motion mode. We compared the signal stability and signal-to-noise ratio (SNR) of each sequence according to motion state and used the coefficients of variation (CoV) to determine the degree of signal stability.

Results

With motion, CAIPIRINHA-VIBE showed the best image quality, and the motion correction aligned the images very well. The CoV (%) of CAIPIRINHA-VIBE in the moving mode (18.65) decreased significantly after the motion correction (2.56) (p < 0.001). In contrast, c-VIBE showed severe breathing motion artifacts that did not improve after motion correction. For radial-VIBE, the position of the phantom in the images did not change during motion, but streak artifacts significantly degraded image quality, also after motion correction. In addition, SNR increased in both CAIPIRINHA-VIBE (from 3.37 to 9.41, p < 0.001) and radial-VIBE (from 4.3 to 4.96, p < 0.001) after motion correction.

Conclusion

CAIPIRINHA-VIBE performed best for free-breathing DCE-MRI after motion correction, with excellent image quality.

SUBMITTER: Lee CK 

PROVIDER: S-EPMC5313517 | biostudies-literature | 2017 Mar-Apr

REPOSITORIES: biostudies-literature

altmetric image

Publications

The Effects of Breathing Motion on DCE-MRI Images: Phantom Studies Simulating Respiratory Motion to Compare CAIPIRINHA-VIBE, Radial-VIBE, and Conventional VIBE.

Lee Chang Kyung CK   Seo Nieun N   Kim Bohyun B   Huh Jimi J   Kim Jeong Kon JK   Lee Seung Soo SS   Kim In Seong IS   Nickel Dominik D   Kim Kyung Won KW  

Korean journal of radiology 20170207 2


<h4>Objective</h4>To compare the breathing effects on dynamic contrast-enhanced (DCE)-MRI between controlled aliasing in parallel imaging results in higher acceleration (CAIPIRINHA)-volumetric interpolated breath-hold examination (VIBE), radial VIBE with k-space-weighted image contrast view-sharing (radial-VIBE), and conventional VIBE (c-VIBE) sequences using a dedicated phantom experiment.<h4>Materials and methods</h4>We developed a moving platform to simulate breathing motion. We conducted dyn  ...[more]

Similar Datasets

| S-EPMC9278689 | biostudies-literature
| S-EPMC6411681 | biostudies-literature
| S-EPMC5646228 | biostudies-literature
| S-EPMC3360690 | biostudies-other
| S-EPMC8024254 | biostudies-literature
| S-EPMC4453263 | biostudies-literature
| S-EPMC5871341 | biostudies-literature
| S-EPMC4409628 | biostudies-literature
| S-EPMC6080614 | biostudies-literature
| S-EPMC5605287 | biostudies-literature