Validation of Contact-Free Sleep Monitoring Device with Comparison to Polysomnography.


ABSTRACT:

Study objectives

To validate a contact-free system designed to achieve maximal comfort during long-term sleep monitoring, together with high monitoring accuracy.

Methods

We used a contact-free monitoring system (EarlySense, Ltd., Israel), comprising an under-the-mattress piezoelectric sensor and a smartphone application, to collect vital signs and analyze sleep. Heart rate (HR), respiratory rate (RR), body movement, and calculated sleep-related parameters from the EarlySense (ES) sensor were compared to data simultaneously generated by the gold standard, polysomnography (PSG). Subjects in the sleep laboratory underwent overnight technician-attended full PSG, whereas subjects at home were recorded for 1 to 3 nights with portable partial PSG devices. Data were compared epoch by epoch.

Results

A total of 63 subjects (85 nights) were recorded under a variety of sleep conditions. Compared to PSG, the contact-free system showed similar values for average total sleep time (TST), % wake, % rapid eye movement, and % non-rapid eye movement sleep, with 96.1% and 93.3% accuracy of continuous measurement of HR and RR, respectively. We found a linear correlation between TST measured by the sensor and TST determined by PSG, with a coefficient of 0.98 (R = 0.87). Epoch-by-epoch comparison with PSG in the sleep laboratory setting revealed that the system showed sleep detection sensitivity, specificity, and accuracy of 92.5%, 80.4%, and 90.5%, respectively.

Conclusions

TST estimates with the contact-free sleep monitoring system were closely correlated with the gold-standard reference. This system shows good sleep staging capability with improved performance over accelerometer-based apps, and collects additional physiological information on heart rate and respiratory rate.

SUBMITTER: Tal A 

PROVIDER: S-EPMC5337599 | biostudies-literature | 2017 Mar

REPOSITORIES: biostudies-literature

altmetric image

Publications

Validation of Contact-Free Sleep Monitoring Device with Comparison to Polysomnography.

Tal Asher A   Shinar Zvika Z   Shaki David D   Codish Shlomi S   Goldbart Aviv A  

Journal of clinical sleep medicine : JCSM : official publication of the American Academy of Sleep Medicine 20170315 3


<h4>Study objectives</h4>To validate a contact-free system designed to achieve maximal comfort during long-term sleep monitoring, together with high monitoring accuracy.<h4>Methods</h4>We used a contact-free monitoring system (EarlySense, Ltd., Israel), comprising an under-the-mattress piezoelectric sensor and a smartphone application, to collect vital signs and analyze sleep. Heart rate (HR), respiratory rate (RR), body movement, and calculated sleep-related parameters from the EarlySense (ES)  ...[more]

Similar Datasets

| S-EPMC6760396 | biostudies-literature
| S-EPMC10238963 | biostudies-literature
| S-EPMC5889179 | biostudies-literature
| S-EPMC9584568 | biostudies-literature
| S-EPMC5135638 | biostudies-literature
| S-EPMC3134551 | biostudies-literature
| S-EPMC10382437 | biostudies-literature
| S-EPMC9578631 | biostudies-literature
| S-EPMC7439146 | biostudies-literature
| S-EPMC7796293 | biostudies-literature