Diagnostic accuracy and added value of dual-energy subtraction radiography compared to standard conventional radiography using computed tomography as standard of reference.


ABSTRACT:

Purpose

To retrospectively evaluate diagnostic performance of dual-energy subtraction radiography (DESR) for interpretation of chest radiographs compared to conventional radiography (CR) using computed tomography (CT) as standard of reference.

Material and methods

A total of 199 patients (75 female, median age 67) were included in this institutional review board (IRB)-approved clinical trial. All patients were scanned in posteroanterior and lateral direction with dual-shot DE-technique. Chest CT was performed within ±72 hours. The system provides three types of images: bone weighted-image, soft tissue weighted-image, herein termed as DESR-images, and a standard image, termed CR-image (marked as CR-image). Images were evaluated by two radiologists for presence of inserted life support lines, pneumothorax, pleural effusion, infectious consolidation, interstitial lung changes, tumor, skeletal alterations, soft tissue alterations, aortic or tracheal calcification and pleural thickening. Inter-observer agreement between readers and diagnostic performance were calculated. McNemar's test was used to test for significant differences.

Results

Mean inter-observer agreement throughout the investigated parameters was higher in DESR images compared to CR-images (kDESR = 0.935 vs. kCR = 0.858). DESR images provided significantly increased sensitivity compared to CR-images for the detection of infectious consolidations (42% vs. 62%), tumor (46% vs. 57%), interstitial lung changes (69% vs. 87%) and aortic or tracheal calcification (25 vs. 73%) (p<0.05). There were no significant differences in sensitivity for the detection of inserted life support lines, pneumothorax, pleural effusion, skeletal alterations, soft tissue alterations or pleural thickening (p>0.05).

Conclusion

DESR increases significantly the sensibility without affecting the specificity evaluating chest radiographs, with emphasis on the detection of interstitial lung diseases.

SUBMITTER: Martini K 

PROVIDER: S-EPMC5354458 | biostudies-literature | 2017

REPOSITORIES: biostudies-literature

altmetric image

Publications

Diagnostic accuracy and added value of dual-energy subtraction radiography compared to standard conventional radiography using computed tomography as standard of reference.

Martini Katharina K   Baessler Marco M   Baumueller Stephan S   Frauenfelder Thomas T  

PloS one 20170316 3


<h4>Purpose</h4>To retrospectively evaluate diagnostic performance of dual-energy subtraction radiography (DESR) for interpretation of chest radiographs compared to conventional radiography (CR) using computed tomography (CT) as standard of reference.<h4>Material and methods</h4>A total of 199 patients (75 female, median age 67) were included in this institutional review board (IRB)-approved clinical trial. All patients were scanned in posteroanterior and lateral direction with dual-shot DE-tech  ...[more]

Similar Datasets

| S-EPMC5082876 | biostudies-literature
| S-EPMC9529964 | biostudies-literature
| S-EPMC5875487 | biostudies-literature
| S-EPMC7668337 | biostudies-literature
| S-EPMC10558368 | biostudies-literature
| S-EPMC6746727 | biostudies-literature
| S-EPMC5270592 | biostudies-literature
| S-EPMC9689116 | biostudies-literature
| S-EPMC5549549 | biostudies-other
| S-EPMC7054418 | biostudies-literature