Estimating scatter from sparsely measured primary signal.


ABSTRACT: Scatter radiation severely degrades the image quality. Measurement-based scatter correction methods sample the scatter signal at sparsely distributed points, from which the scatter profile is estimated and deterministically removed from the projection image. The estimation of the scatter profile is generally done through a spline interpolation and the resulting scatter profile is quite smooth. Consequently, the noise is intact and the signal-to-noise ratio is reduced in the projection image after scatter correction, leading to image artifacts and increased noise in the reconstruction images. We propose a simple and effective method, referred to as filtered scatter-to-primary ratio ([Formula: see text]-SPR) estimation, to estimate the scatter profile using the sparsely sampled scatter signal. Using the primary sampling device and the stationary digital tomosynthesis systems previously developed in our lab, we evaluated and compared the [Formula: see text]-SPR method in comparison with existing methods in terms of contrast ratio, signal difference-to-noise ratio, and modulation transfer function. A significant improvement in image quality is observed in both the projection and the reconstruction images using the proposed method.

SUBMITTER: Wu G 

PROVIDER: S-EPMC5370239 | biostudies-literature | 2017 Jan

REPOSITORIES: biostudies-literature

altmetric image

Publications

Estimating scatter from sparsely measured primary signal.

Wu Gongting G   Inscoe Christina R CR   Calliste Jabari J   Shan Jing J   Lee Yueh Z YZ   Zhou Otto O   Lu Jianping J  

Journal of medical imaging (Bellingham, Wash.) 20170101 1


Scatter radiation severely degrades the image quality. Measurement-based scatter correction methods sample the scatter signal at sparsely distributed points, from which the scatter profile is estimated and deterministically removed from the projection image. The estimation of the scatter profile is generally done through a spline interpolation and the resulting scatter profile is quite smooth. Consequently, the noise is intact and the signal-to-noise ratio is reduced in the projection image afte  ...[more]

Similar Datasets

| S-EPMC2658051 | biostudies-literature
| S-EPMC9831746 | biostudies-literature
| S-EPMC8596391 | biostudies-literature
| S-EPMC7673459 | biostudies-literature
| S-EPMC8320731 | biostudies-literature
| S-EPMC5748468 | biostudies-literature
2019-10-28 | GSE125049 | GEO
| S-EPMC6075374 | biostudies-literature
| S-EPMC4247041 | biostudies-literature
| S-EPMC7769416 | biostudies-literature