Exploration of the Performance of a Hybrid Closed Loop Insulin Delivery Algorithm That Includes Insulin Delivery Limits Designed to Protect Against Hypoglycemia.


ABSTRACT:

Background

Hypoglycemia remains a risk for closed loop insulin delivery particularly following exercise or if the glucose sensor is inaccurate. The aim of this study was to test whether an algorithm that includes a limit to insulin delivery is effective at protecting against hypoglycemia under those circumstances.

Methods

An observational study on 8 participants with type 1 diabetes was conducted, where a hybrid closed loop system (HCL) (Medtronic™ 670G) was challenged with hypoglycemic stimuli: exercise and an overreading glucose sensor.

Results

There was no overnight or exercise-induced hypoglycemia during HCL insulin delivery. All daytime hypoglycemia was attributable to postmeal bolused insulin in those participants with a more aggressive carbohydrate factor.

Conclusion

HCL systems rely on accurate carbohydrate ratios and carbohydrate counting to avoid hypoglycemia. The algorithm that was tested against moderate exercise and an overreading glucose sensor performed well in terms of hypoglycemia avoidance. Algorithm refinement continues in preparation for long-term outpatient trials.

SUBMITTER: de Bock M 

PROVIDER: S-EPMC5375079 | biostudies-literature | 2017 Jan

REPOSITORIES: biostudies-literature

altmetric image

Publications

Exploration of the Performance of a Hybrid Closed Loop Insulin Delivery Algorithm That Includes Insulin Delivery Limits Designed to Protect Against Hypoglycemia.

de Bock Martin M   Dart Julie J   Roy Anirban A   Davey Raymond R   Soon Wayne W   Berthold Carolyn C   Retterath Adam A   Grosman Benyamin B   Kurtz Natalie N   Davis Elizabeth E   Jones Timothy T  

Journal of diabetes science and technology 20160925 1


<h4>Background</h4>Hypoglycemia remains a risk for closed loop insulin delivery particularly following exercise or if the glucose sensor is inaccurate. The aim of this study was to test whether an algorithm that includes a limit to insulin delivery is effective at protecting against hypoglycemia under those circumstances.<h4>Methods</h4>An observational study on 8 participants with type 1 diabetes was conducted, where a hybrid closed loop system (HCL) (Medtronic™ 670G) was challenged with hypogl  ...[more]

Similar Datasets

| S-EPMC9464083 | biostudies-literature
| S-EPMC5769873 | biostudies-literature
| S-EPMC8792215 | biostudies-literature
| S-EPMC7896863 | biostudies-literature
| S-EPMC3781513 | biostudies-literature
| S-EPMC9983122 | biostudies-literature
| S-EPMC6609966 | biostudies-literature
| S-EPMC10126177 | biostudies-literature
| S-EPMC4183919 | biostudies-literature
| S-EPMC10698209 | biostudies-literature