Influence of eye biometrics and corneal micro-structure on noncontact tonometry.


ABSTRACT:

Purpose

Tonometry is widely used as the main screening tool supporting glaucoma diagnosis. Still, its accuracy could be improved if full knowledge about the variation of the corneal biomechanical properties was available. In this study, Optical Coherence Tomography (OCT) speckle statistics are used to infer the organisation of the corneal micro-structure and hence, to analyse its influence on intraocular pressure (IOP) measurements.

Methods

Fifty-six subjects were recruited for this prospective study. Macro and micro-structural corneal parameters as well as subject age were considered. Macro-structural analysis included the parameters that are associated with the ocular anatomy, such as central corneal thickness (CCT), corneal radius, axial length, anterior chamber depth and white-to-white corneal diameter. Micro-structural parameters which included OCT speckle statistics were related to the internal organisation of the corneal tissue and its physiological changes during lifetime. The corneal speckle obtained from OCT was modelled with the Generalised Gamma (GG) distribution that is characterised with a scale parameter and two shape parameters.

Results

In macro-structure analysis, only CCT showed a statistically significant correlation with IOP (R2 = 0.25, p<0.001). The scale parameter and the ratio of the shape parameters of GG distribution showed statistically significant correlation with IOP (R2 = 0.19, p<0.001 and R2 = 0.17, p<0.001, respectively). For the studied group, a weak, although significant correlation was found between age and IOP (R2 = 0.053, p = 0.04). Forward stepwise regression showed that CCT and the scale parameter of the Generalised Gamma distribution can be combined in a regression model (R2 = 0.39, p<0.001) to study the role of the corneal structure on IOP.

Conclusions

We show, for the first time, that corneal micro-structure influences the IOP measurements obtained from noncontact tonometry. OCT speckle statistics can be employed to learn about the corneal micro-structure and hence, to further calibrate the IOP measurements.

SUBMITTER: Jesus DA 

PROVIDER: S-EPMC5417684 | biostudies-literature | 2017

REPOSITORIES: biostudies-literature

altmetric image

Publications

Influence of eye biometrics and corneal micro-structure on noncontact tonometry.

Jesus Danilo A DA   Majewska Małgorzata M   Krzyżanowska-Berkowska Patrycja P   Iskander D Robert DR  

PloS one 20170504 5


<h4>Purpose</h4>Tonometry is widely used as the main screening tool supporting glaucoma diagnosis. Still, its accuracy could be improved if full knowledge about the variation of the corneal biomechanical properties was available. In this study, Optical Coherence Tomography (OCT) speckle statistics are used to infer the organisation of the corneal micro-structure and hence, to analyse its influence on intraocular pressure (IOP) measurements.<h4>Methods</h4>Fifty-six subjects were recruited for th  ...[more]