Respiratory motion-resolved, self-gated 4D-MRI using rotating cartesian k-space (ROCK).


ABSTRACT: To propose and validate a respiratory motion resolved, self-gated (SG) 4D-MRI technique to assess patient-specific breathing motion of abdominal organs for radiation treatment planning.The proposed 4D-MRI technique was based on the balanced steady-state free-precession (bSSFP) technique and 3D k-space encoding. A novel rotating cartesian k-space (ROCK) reordering method was designed which incorporates repeatedly sampled k-space centerline as the SG motion surrogate and allows for retrospective k-space data binning into different respiratory positions based on the amplitude of the surrogate. The multiple respiratory-resolved 3D k-space data were subsequently reconstructed using a joint parallel imaging and compressed sensing method with spatial and temporal regularization. The proposed 4D-MRI technique was validated using a custom-made dynamic motion phantom and was tested in six healthy volunteers, in whom quantitative diaphragm and kidney motion measurements based on 4D-MRI images were compared with those based on 2D-CINE images.The 5-minute 4D-MRI scan offers high-quality volumetric images in 1.2 × 1.2 × 1.6 mm3 and eight respiratory positions, with good soft-tissue contrast. In phantom experiments with triangular motion waveform, the motion amplitude measurements based on 4D-MRI were 11.89% smaller than the ground truth, whereas a -12.5% difference was expected due to data binning effects. In healthy volunteers, the difference between the measurements based on 4D-MRI and the ones based on 2D-CINE were 6.2 ± 4.5% for the diaphragm, 8.2 ± 4.9% and 8.9 ± 5.1% for the right and left kidney.The proposed 4D-MRI technique could provide high-resolution, high-quality, respiratory motion-resolved 4D images with good soft-tissue contrast and are free of the "stitching" artifacts usually seen on 4D-CT and 4D-MRI based on resorting 2D-CINE. It could be used to visualize and quantify abdominal organ motion for MRI-based radiation treatment planning.

SUBMITTER: Han F 

PROVIDER: S-EPMC5462449 | biostudies-literature | 2017 Apr

REPOSITORIES: biostudies-literature

altmetric image

Publications

Respiratory motion-resolved, self-gated 4D-MRI using rotating cartesian k-space (ROCK).

Han Fei F   Zhou Ziwu Z   Cao Minsong M   Yang Yingli Y   Sheng Ke K   Hu Peng P  

Medical physics 20170311 4


<h4>Purpose</h4>To propose and validate a respiratory motion resolved, self-gated (SG) 4D-MRI technique to assess patient-specific breathing motion of abdominal organs for radiation treatment planning.<h4>Methods</h4>The proposed 4D-MRI technique was based on the balanced steady-state free-precession (bSSFP) technique and 3D k-space encoding. A novel rotating cartesian k-space (ROCK) reordering method was designed which incorporates repeatedly sampled k-space centerline as the SG motion surrogat  ...[more]

Similar Datasets

| S-EPMC5313377 | biostudies-literature
| S-EPMC6220806 | biostudies-literature
| S-EPMC8172093 | biostudies-literature
| S-EPMC6330120 | biostudies-literature
| S-EPMC7755133 | biostudies-literature
| S-EPMC8634777 | biostudies-literature
| S-EPMC10149606 | biostudies-literature
| S-EPMC7285911 | biostudies-literature
| S-EPMC8627452 | biostudies-literature
| S-EPMC7027768 | biostudies-literature