Native MS Analysis of Bacteriorhodopsin and an Empty Nanodisc by Orthogonal Acceleration Time-of-Flight, Orbitrap and Ion Cyclotron Resonance.


ABSTRACT: Over the past two decades, orthogonal acceleration time-of-flight has been the de facto analyzer for solution and membrane-soluble protein native mass spectrometry (MS) studies; this however is gradually changing. Three MS instruments are compared, the Q-ToF, Orbitrap, and the FT-ICR, to analyze, under native instrument and buffer conditions, the seven-transmembrane helical protein bacteriorhodopsin-octylglucoside micelle and the empty nanodisc (MSP1D1-Nd) using both MS and tandem-MS modes of operation. Bacteriorhodopsin can be released from the octylglucoside-micelle efficiently on all three instruments (MS-mode), producing a narrow charge state distribution (z = 8+ to 10+) by either increasing the source lens or collision cell (or HCD) voltages. A lower center-of-mass collision energy (0