The combination of the error correction methods of GAFCHROMIC EBT3 film.


ABSTRACT:

Purpose

The aim of this study was to combine a set of methods for use of radiochromic film dosimetry, including calibration, correction for lateral effects and a proposed triple-channel analysis. These methods can be applied to GAFCHROMIC EBT3 film dosimetry for radiation field analysis and verification of IMRT plans.

Methods

A single-film exposure was used to achieve dose calibration, and the accuracy was verified based on comparisons with the square-field calibration method. Before performing the dose analysis, the lateral effects on pixel values were corrected. The position dependence of the lateral effect was fitted by a parabolic function, and the curvature factors of different dose levels were obtained using a quadratic formula. After lateral effect correction, a triple-channel analysis was used to reduce disturbances and convert scanned images from films into dose maps. The dose profiles of open fields were measured using EBT3 films and compared with the data obtained using an ionization chamber. Eighteen IMRT plans with different field sizes were measured and verified with EBT3 films, applying our methods, and compared to TPS dose maps, to check correct implementation of film dosimetry proposed here.

Results

The uncertainty of lateral effects can be reduced to ±1 cGy. Compared with the results of Micke A et al., the residual disturbances of the proposed triple-channel method at 48, 176 and 415 cGy are 5.3%, 20.9% and 31.4% smaller, respectively. Compared with the ionization chamber results, the difference in the off-axis ratio and percentage depth dose are within 1% and 2%, respectively. For the application of IMRT verification, there were no difference between two triple-channel methods. Compared with only corrected by triple-channel method, the IMRT results of the combined method (include lateral effect correction and our present triple-channel method) show a 2% improvement for large IMRT fields with the criteria 3%/3 mm.

SUBMITTER: Li Y 

PROVIDER: S-EPMC5531657 | biostudies-literature | 2017

REPOSITORIES: biostudies-literature

altmetric image

Publications

The combination of the error correction methods of GAFCHROMIC EBT3 film.

Li Yinghui Y   Chen Lixin L   Zhu Jinhan J   Liu Xiaowei X  

PloS one 20170727 7


<h4>Purpose</h4>The aim of this study was to combine a set of methods for use of radiochromic film dosimetry, including calibration, correction for lateral effects and a proposed triple-channel analysis. These methods can be applied to GAFCHROMIC EBT3 film dosimetry for radiation field analysis and verification of IMRT plans.<h4>Methods</h4>A single-film exposure was used to achieve dose calibration, and the accuracy was verified based on comparisons with the square-field calibration method. Bef  ...[more]

Similar Datasets

| S-EPMC5690916 | biostudies-literature
| S-EPMC6362602 | biostudies-literature
| S-EPMC7751105 | biostudies-literature
| S-EPMC7079412 | biostudies-literature
| S-EPMC6966875 | biostudies-literature
| S-EPMC4276494 | biostudies-literature
| S-EPMC8632833 | biostudies-literature
| S-EPMC7517614 | biostudies-literature
2023-05-11 | GSE232200 | GEO
| S-EPMC5271325 | biostudies-literature