Unknown

Dataset Information

0

Adenosine triphosphate bioluminescence for bacteriologic surveillance and reprocessing strategies for minimizing risk of infection transmission by duodenoscopes.


ABSTRACT: BACKGROUND AND AIMS:Recent outbreaks of duodenoscope-transmitted infections underscore the importance of adequate endoscope reprocessing. Adenosine triphosphate (ATP) bioluminescence testing allows rapid evaluation of endoscopes for bacteriologic/biologic residue. In this prospective study we evaluate the utility of ATP in bacteriologic surveillance and the effects of endoscopy staff education and dual cycles of cleaning and high-level disinfection (HLD) on endoscope reprocessing. METHODS:ATP bioluminescence was measured after precleaning, manual cleaning, and HLD on rinsates from suction-biopsy channels of all endoscopes and elevator channels of duodenoscopes/linear echoendoscopes after use. ATP bioluminescence was remeasured in duodenoscopes (1) after re-education and competency testing of endoscopy staff and subsequently (2) after 2 cycles of precleaning and manual cleaning and single cycle of HLD or (3) after 2 cycles of precleaning, manual cleaning, and HLD. RESULTS:The ideal ATP bioluminescence benchmark of <200 relative light units (RLUs) after manual cleaning was achieved from suction-biopsy channel rinsates of all endoscopes, but 9 of 10 duodenoscope elevator channel rinsates failed to meet this benchmark. Re-education reduced RLUs in duodenoscope elevator channel rinsates after precleaning (23,218.0 vs 1340.5 RLUs, P < .01) and HLD (177.0 vs 12.0 RLUs, P < .01). After 2 cycles of manual cleaning/HLD, duodenoscope elevator channel RLUs achieved levels similar to sterile water, with corresponding negative cultures. CONCLUSIONS:ATP testing offers a rapid, inexpensive alternative for detection of endoscope microbial residue. Re-education of endoscopy staff and 2 cycles of cleaning and HLD decreased elevator channel RLUs to levels similar to sterile water and may therefore minimize the risk of transmission of infections by duodenoscopes.

SUBMITTER: Sethi S 

PROVIDER: S-EPMC5534172 | biostudies-literature | 2017 Jun

REPOSITORIES: biostudies-literature

altmetric image

Publications

Adenosine triphosphate bioluminescence for bacteriologic surveillance and reprocessing strategies for minimizing risk of infection transmission by duodenoscopes.

Sethi Saurabh S   Huang Robert J RJ   Barakat Monique T MT   Banaei Niaz N   Friedland Shai S   Banerjee Subhas S  

Gastrointestinal endoscopy 20161103 6


<h4>Background and aims</h4>Recent outbreaks of duodenoscope-transmitted infections underscore the importance of adequate endoscope reprocessing. Adenosine triphosphate (ATP) bioluminescence testing allows rapid evaluation of endoscopes for bacteriologic/biologic residue. In this prospective study we evaluate the utility of ATP in bacteriologic surveillance and the effects of endoscopy staff education and dual cycles of cleaning and high-level disinfection (HLD) on endoscope reprocessing.<h4>Met  ...[more]

Similar Datasets

| S-EPMC6754731 | biostudies-literature
| S-EPMC6786574 | biostudies-literature
| S-EPMC7409044 | biostudies-literature
| S-EPMC6693643 | biostudies-literature
| S-EPMC6885784 | biostudies-literature
| S-EPMC2744016 | biostudies-literature
| S-EPMC6642494 | biostudies-literature
| S-EPMC6044877 | biostudies-literature
| S-EPMC5929190 | biostudies-literature
| S-EPMC6248038 | biostudies-literature