ABSTRACT: Objective
A fully closed-loop insulin-only system was developed to provide glucose control in patients with type 1 diabetes without requiring announcement of meals or activity. Our goal was to assess initial safety and efficacy of this system.Research design and methods
The multiple model probabilistic controller (MMPPC) anticipates meals when the patient is awake. The controller used the subject's basal rates and total daily insulin dose for initialization. The system was tested at two sites on 10 patients in a 30-h inpatient study, followed by 15 subjects at three sites in a 54-h supervised hotel study, where the controller was challenged by exercise and unannounced meals. The system was implemented on the UVA DiAs system using a Roche Spirit Combo Insulin Pump and a Dexcom G4 Continuous Glucose Monitor.Results
The mean overall (24-h basis) and nighttime (11 PM-7 AM) continuous glucose monitoring (CGM) values were 142 and 125?mg/dL during the inpatient study. The hotel study used a different daytime tuning and manual announcement, instead of automatic detection, of sleep and wake periods. This resulted in mean overall (24-h basis) and nighttime CGM values of 152 and 139?mg/dL for the hotel study and there was also a reduction in hypoglycemia events from 1.6 to 0.91 events/patient/day.Conclusions
The MMPPC system achieved a mean glucose that would be particularly helpful for people with an elevated A1c as a result of frequent missed meal boluses. Current full closed loop has a higher risk for hypoglycemia when compared with algorithms using meal announcement.
SUBMITTER: Cameron FM
PROVIDER: S-EPMC5647490 | biostudies-literature | 2017 Sep
REPOSITORIES: biostudies-literature
Cameron Faye M FM Ly Trang T TT Buckingham Bruce A BA Maahs David M DM Forlenza Gregory P GP Levy Carol J CJ Lam David D Clinton Paula P Messer Laurel H LH Westfall Emily E Levister Camilla C Xie Yan Yan YY Baysal Nihat N Howsmon Daniel D Patek Stephen D SD Bequette B Wayne BW
Diabetes technology & therapeutics 20170802 9
<h4>Objective</h4>A fully closed-loop insulin-only system was developed to provide glucose control in patients with type 1 diabetes without requiring announcement of meals or activity. Our goal was to assess initial safety and efficacy of this system.<h4>Research design and methods</h4>The multiple model probabilistic controller (MMPPC) anticipates meals when the patient is awake. The controller used the subject's basal rates and total daily insulin dose for initialization. The system was tested ...[more]