ABSTRACT: Background
Allospecific anti-HLA antibodies (Abs) are associated with rejection of solid organ grafts. The 2 main kits to detect anti-HLA Ab in patient serum are commercialized by Immucor and One Lambda/ThermoFisher. We sought to compare the performance of both platforms.Methods
Background-adjusted mean fluorescence intensity (MFI) values were used from both platforms to compare sera collected from 125 pretransplant and posttransplant heart and lung transplant recipients.Results
Most HLA class I (94.5%) and HLA class II (89%) Abs with moderate to high MFI titer (≥4000) were detected by both assays. A modest correlation was observed between MFI values obtained from the 2 assays for both class I (r = 0.3, r2 = 0.09, P < 0.0001) and class II Ab (r = 0.707, r2 = 0.5, P < 0.0001). Both assays detected anti-class I and II Ab that the other did not; however, no specific HLA allele was detected preferentially by either of the 2 assays. For a limited number of discrepant sera, dilution resulted in comparable reactivity profiles between the 2 platforms.Conclusions
Immucor and One Lambda/ThermoFisher assays have a similar, albeit nonidentical, ability to detect anti-HLA Ab. Although the correlation between the assays was present, significant variances exist, some of which can be explained by a dilution-sensitive "prozone" effect.
SUBMITTER: Clerkin KJ
PROVIDER: S-EPMC5682763 | biostudies-literature | 2017 Nov
REPOSITORIES: biostudies-literature
Clerkin Kevin J KJ See Sarah B SB Farr Maryjane A MA Restaino Susan W SW Serban Geo G Latif Farhana F Li Lingzhi L Colombo Paolo C PC Vlad George G Ray Bryan B Vasilescu Elena R ER Zorn Emmanuel E
Transplantation direct 20171002 11
<h4>Background</h4>Allospecific anti-HLA antibodies (Abs) are associated with rejection of solid organ grafts. The 2 main kits to detect anti-HLA Ab in patient serum are commercialized by Immucor and One Lambda/ThermoFisher. We sought to compare the performance of both platforms.<h4>Methods</h4>Background-adjusted mean fluorescence intensity (MFI) values were used from both platforms to compare sera collected from 125 pretransplant and posttransplant heart and lung transplant recipients.<h4>Resu ...[more]